![YOKOGAWA YTA110 User Manual Download Page 3](http://html1.mh-extra.com/html/yokogawa/yta110/yta110_user-manual_3835920003.webp)
3
2.4 KS25 pending the sale to EEA market.
Item
Description
Code
ATEX
ATEX Intrinsically Safe “ia” Approval
Electrical Connection: 1/2 NPT female and M20 female
KS25
(Note1)
Note1: Because the examination of conformity to the updated standards is under the way, intrinsically safe type Ex ia
(Option: KS25 for Fieldbus communication type) is pending the sale to EEA market.
2.5 Change Name Plate.
2.5.1 IM 01C50B01-01E
MODEL
NO.
OUTPUT
CAL
RNG
SUFFIX
STYLE
SUPPLY
II 1 G
No. KEMA 02ATEX1026 X
Ex ia IIC T5 Ga
Tamb -40 TO 50°C
Ex ia IIC T4 Ga
Tamb -40 TO 70°C
ENCLOSURE: IP66/IP67
SUPPLY INPUT
Ui=30V, Ii=165mA, Pi=900mW
Ci=20nF, Li=730µH
SENSOR OUTPUT
Uo=9V, Io=40mA, Po=90mW
Co=0.7µF, Lo=10mH
POTENTIAL ELECTROSTATIC
CHARGING HAZARD
-SEE USER’S MANUAL
II 3 G
Ex ic IIC T5...T4 Gc
Tamb -30 TO 70°C
(50°C for T5)
ENCLOSURE: IP66/IP67
SUPPLY INPUT
Ui=30V
Ci=28nF, Li=730µH
SENSOR OUTPUT
Uo=9V, Io=40mA, Po=90mW
Co=0.7µF, Lo=10mH
II 2 GD
No. KEMA 07ATEX0130
Ex d IIC T6/T5 Gb
Ex tb IIIC T70°C, T90°C Db
TEMP. CLASS T6/T5
Tamb(Gas) -40 to +75°C(T6)
-40 to +80°C(T5)
(Dust) -30 to +65°C(T70°C)
-30 to +80°C(T90°C)
ENCLOSURE: IP66/IP67
AFTER DE-ENERGIZING, DELAY
5 MINUTES BEFORE OPENING.
WHEN THE AMBIENT TEMP.≥70°C,
USE THE HEAT-RESISTING
CABLES & CABLE GLANDS≥90°C.
POTENTIAL ELECTROSTATIC
CHARGING HAZARD
-SEE USER’S MANUAL
POTENTIAL ELECTROSTATIC
CHARGING HAZARD
-SEE USER’S MANUAL
II 1 G
No. KEMA 02ATEX1026 X
Ex ia IIC T5 Ga
Tamb -40 TO 50°C
Ex ia IIC T4 Ga
Tamb -40 TO 70°C
ENCLOSURE: IP66/IP67
SUPPLY INPUT
Ui=30V, Ii=165mA, Pi=900mW
Ci=20nF, Li=730µH
SENSOR OUTPUT
Uo=9V, Io=40mA, Po=90mW
Co=0.7µF, Lo=10mH
POTENTIAL ELECTROSTATIC
CHARGING HAZARD
-SEE USER’S MANUAL
II 3 G
Ex ic IIC T5...T4 Gc
Tamb -30 TO 70°C
(50°C for T5)
ENCLOSURE: IP66/IP67
SUPPLY INPUT
Ui=30V
Ci=28nF, Li=730µH
SENSOR OUTPUT
Uo=9V, Io=40mA, Po=90mW
Co=0.7µF, Lo=10mH
II 2 GD
No. KEMA 07ATEX0130
Ex db IIC T6/T5 Gb
Ex tb IIIC T70°C/T90°C Db
TEMP. CLASS T6/T5
Tamb(Gas) -40 to +75°C(T6)
-40 to +80°C(T5)
(Dust) -30 to +65°C(T70°C)
-30 to +80°C(T90°C)
ENCLOSURE: IP66/IP67
AFTER DE-ENERGIZING, DELAY
5 MINUTES BEFORE OPENING.
WHEN THE AMBIENT TEMP.≥70°C,
USE THE HEAT-RESISTING
CABLES & CABLE GLANDS≥90°C.
POTENTIAL ELECTROSTATIC
CHARGING HAZARD
-SEE USER’S MANUAL
POTENTIAL ELECTROSTATIC
CHARGING HAZARD
-SEE USER’S MANUAL
Before Change
After Change
MODEL
NO.
OUTPUT
CAL
RNG
SUFFIX
STYLE
SUPPLY
2.5.2 IM 01C50T02-01E
Name plate for flameproof type
Name plate for flameproof type
Before Change
After Change
TEMPERATURE
TRANSMITTER
YTA
MODEL
SUFFIX
STYLE
SUPPLY
NO.
OUTPUT
CAL
RNG
TEMPERATURE
TRANSMITTER
YTA
MODEL
SUFFIX
STYLE
SUPPLY
NO.
OUTPUT
CAL
RNG
No. KEMA 07ATEX0130
Ex d IIC T6/T5 Gb
Ex tb IIIC T70°C, T90°C Db
TEMP. CLASS T6/T5
Tamb (Gas) -40 to +75°C(T6)
-40 to +80°C(T5)
(Dust) -30 to +65°C(T70°C)
-30 to +80°C(T90°C)
ENCLOSURE: IP66/IP67
WARNING
AFTER DE-ENERGIZING. DELAY
5 MINUTES BEFORE OPENING.
WHEN THE AMBIENT TEMP.≥70°C,
USE THE HEAT-RESISTING
CABLES & CABLE GLANDS≥90°C.
POTENTIAL ELECTROSTATIC
CHARGING HAZARD
-SEE USER’ S MANUAL
WARNING
AFTER DE-ENERGIZING. DELAY
5 MINUTES BEFORE OPENING.
WHEN THE AMBIENT TEMP.≥70°C,
USE THE HEAT-RESISTING
CABLES & CABLE GLANDS≥90°C.
POTENTIAL ELECTROSTATIC
CHARGING HAZARD
-SEE USER’ S MANUAL
No. KEMA 07ATEX0130
Ex db IIC T6/T5 Gb
Ex tb IIIC T70°C/T90°C Db
TEMP. CLASS T6/T5
Tamb (Gas) -40 to +75°C(T6)
-40 to +80°C(T5)
(Dust) -30 to +65°C(T70°C)
-30 to +80°C(T90°C)
ENCLOSURE: IP66/IP67
Manual Change No. 16-007-2