Ungrounded Measurement Circuits
When the circuit under measurement is not grounded and the signal is high in frequency or large
in power, the effects of the inductance of the shunt-type current sensor cable become large. In this
case, use an isolation sensor (CT, DC-CT, or clamp) to perform measurements.
Clamp-type current sensor
External current sensor
input connector
LOAD
V
± Voltage input terminals
WT310/WT310HC/WT330
Connecting the 96001 Clamp-on Probe
When using the Yokogawa 96001 Clamp-on Probe, use the 758924 Conversion Adapter (optional
accessory).
96001 (voltage output type)
758924
WT310/WT310HC
WT330
EXT input terminal
EXT input terminal