9-2
IM 253421-01E
9.2
Remote Control (optional)
Controlling Integration
To control integration, apply timing signals according to the timing chart below.
EXT START
EXT STOP
EXT RESET
INTEG BUSY
Stop
Start
Reset
Stop
Start
5 ms min.
5 ms min.
5 ms
Approx.
15 ms
Approx.
15 ms
Approx.
15 ms
Approx.
15 ms
As shown in the timing chart, the INTEG BUSY output
signal level goes low while integration is in progress.
The signal can be used to monitor integration, etc.
Holding Display Data Update (same function as HOLD key)
5 ms min.
Display hold
EXT. HOLD
To hold the display update, apply the EXT. HOLD signal according to the timing chart below.
Updating Display Data which has been held (same function as TRIG key)
Measurement start
EXT. TRIG
5 ms min.
250 ms min.
Display update
EXT. TRIG
5 ms min.
Display update
5 ms min.
3 s min.
Measurement start
⋅
Update timing during normal measurement/integration
⋅
Update timing while harmonic analysis function is in progress
Applying an EXT.TRIG signal when the display is on hold updates the display data.
CAUTION
Do not apply a voltage which exceeds the TTL level to the remote controller pin.
Also, do not short the output pins nor apply a voltage to them.
The instrument might be damaged.