12.3 Performing Zero Level Compensation
Keys
DC
SCALING
AVG
FILTER
STORE
RECALL
HARMONICS
M W
h
k
A
h
m V
Hz
M W
k
A
deg
var
TIME
FUNCTION
AUTO
MODE
TRIG
MAX HOLD
HOLD
CAL
INTEGRATOR
HARMONICS
INTERFACE
OUTPUT
SHIFT
REMOTE
MEMORY
INTEG SET
ENTER
A RANGE
V RANGE
START
LOCAL
SETUP
STOP
RESET
AUTO
FUNCTION
FUNCTION
m V
PF
M W
k
A
m V
VA
V MEAN
RMS
A OVER
V OVER
SAMPLE
A
B
C
MODE
Displays
relevant
keys and
indicator
Procedure
SHIFT
ENTER
(Execute zero level compensation)
CAL
Explanation
Zero level compensation
This function is used to create a zero input condition using the internal circuit of the
instrument and setting the level at that point to zero level. Zero level compensation
must be performed to meet the specifications of this instrument (see chapter 16).
• Press the SHIFT key to turn ON the indicator to the left of the SHIFT key and
then press the ENTER key to perform zero level compensation.
• Zero compensation is also performed the first time a measurement is made after
changing the measurement range.
Note
•
For making accurate measurements, we recommend zero level compensation to be
performed after warming up the instrument for at least 30 minutes. In addition, the ambient
temperature need to be within the specifications (see chapter 16) and stable.
•
If the measurement mode, measurement range, and input filter are not changed over a long
period of time, the zero level may change due to the changes in the environment surrounding
the instrument. We recommend zero level compensation to be performed in this case.