background image

User’s

Manual

Disk No. WE07

1st Edition: December 1998 (YK)

All Rights Reserved, Copyright © 1998 Yokogawa Electric Corporation

Model 707031/707032

WE7031/WE7032

Optical Interface Module

This User’s Manual contains information about the installation procedures for the optical interface module WE7031/

WE7032 for the PC-based measurement instrument, WE7000.  For other information about the functions and

specifications of the module, read the WE7000 User’s Manual, IM707001-01E.

Checking the Product

Check that the model name given on the name plate matches that on the order.

MODEL

Model

Description

707031

WE7031 optical interface module for measuring station, 1 port

707032

WE7032 optical interface module for measuring station, 2 ports

NO.

When contacting the dealer from which you purchased the instrument, please quote the

instrument No.

CAUTION

• To avoid damaging the instrument when installing modules, make sure

to turn OFF the main power switch of the measuring station.

• Be careful not to get your fingers caught in the ejection lever while

inserting the module.

Installation Procedure

1.

Verify that the power supply is not connected to the measuring station or that the

main power switch is turned OFF.  Make sure to turn off the main power switch on

the rear panel (see WE7000 User’s Manual IM707001-01E).

2.

Remove the cover plate from slot 0.  If another module is already installed in the slot,

remove that module.

3.

Insert the optical interface module into slot 0.  Press firmly until it is securely

connected to the connector.

4.

Fasten the module at both the top and bottom locations with the M3 attachment

screws.  Set the tightening torque to 0.6 to 0.7 N·m.

When removing the module, unfasten the screws and open the eject lever as shown in

the figure on the lower right.

Eject lever

Eject lever

Cover plate

Slot 0

Module attachment
screw

Module attachment
screw

IM 707031-01E 1st Edition 1/2