10.LIST OF PARAMETERS

No.

                         Item

   Display               Remark         No.

                   Item

Display             Remark

 01       Model                                                             MODEL

 02       Tag no.                                                            TAG NO

 03       Self-check result                                            SELF CHK

Display items

  A        Dispay1                                                          DISPLAY1                                        B         Dsplay2                                                DISPLAY2

A01      Input value                                                     INPUT1                                           B01       Input value                                           INPUT1

A02      Output value                                                  OUTPUT                                         B02       Output value                                        OUTPUT

A07      Alarm-1 status                                               ALM1 STATUS                                B07       Alarm-1 status                                      ALM1 STATUS

A08      Alarm-2 status                                               ALM2 STATUS                                B08       Alarm-2 status                                      ALM2 STATUS

A54      Status                                                            STATUS                     *1                  B60       Self-check result                                   SELF CHK

A56      Rev. no                                                           REV NO

A58      Menu rev.                                                       MENU REV

*1: The Status is displayed for service personnel to see history records.

A60      Self-check result                                            SELF CHK

Setting items

  D        Setting (I/O)                                                   SET(I/O)                                            E        Setting (alarm output)                           SET(ALM)

D01      Tag no. 1                                                         TAG NO.1                                       E03       Alarm-1 setting                                     SET POINT1

D02      Tag no. 2                                                         TAG NO.2                                       E04       Alarm-2 setting                                     SET POINT2

D03      Comment 1                                                    COMMENT1                                   E05       Direction of alarm-1 action                   ALM1 ACTION

D04      Comment 2                                                    COMMENT2                                   E06       Direction of alarm-2 action                   ALM2 ACTION

D10      Output range unit                                           UNIT                                               E09        Alarm-1 hysteresis                               HYSTERESIS1

D14      Low output cut point                                       LOW CUT                                      E10        Alarm-2 hysteresis                               HYSTERESIS2

D16      Input type                                                       INP TYPE                                       E11        Alarm-1 ON delay setting                    ON DELAY1

D17      Selection of input hard range                        SELECT RANGE                            E12        Alarm-2 ON delay setting                    ON DELAY2

D18      Input resistance                                             IN RESIST                                      E13        Alarm-1 OFF delay setting                  OFF DELAY1

D22      Input low range                                              INPUT1 L_RNG                              E14       Alarm-2 OFF delay setting                   OFF DELAY2

D23      Input high range                                             INPUT1 H_RNG                             E15       Direction of alarm-1 relay action          RL1 ACTION

D24      Output low range                                           OUT L_RNG                                   E16        Direction of alarm-2 relay action          RL2 ACTION

D25      Output high range                                          OUT H_RNG                                  E60        Self-check result                                  SELF CHK

D46      Pulse width type                                             PULSE TYPE                                  F           Setting (communication)                      SET(COM)

D47      Pulse width time                                             PULSE WIDTH                              F01         Communication protocol                      PROTOCOL

D60      Self-check result                                            SELF CHK                                      F02        Address                                                ADDRESS

                                                  F03         Baud rate                                             BAUD RATE

                                                  F04         Parity                                                    PARITY

There are items not displayed depending on what output-2 is specified.                          F05         Data length                                           DATA LEN

                                                  F06         Stop bit                                                 STOP BIT

                                                  F07         Decimal point position of input             INPUT DEC PT

                                                  F60         Self-check result                                   SELF CHK

                                      Adjusting items

                                 Test items

  P        Adjustment                                                      ADJUST                                          Q           Test                                                      TEST

P02      Zero adjustment of input-1                              ZERO ADJ                                    Q04         Forced output (alarm)                          ALM1 TEST

P03      Span adjustment of input-1                             SPAN ADJ1                                   Q05         Forced output (alarm)                          ALM2 TEST

P17      Adjustment of external input resistance          RESISTOR ADJ                            Q08         Forced output                                       OUT TEST

P60      Self-check result                                             SELF CHK                                     Q60         Self-check result                                  SELF CHK

11.MAINTENANCE

The product starts running immediately when the power is turned on;
however, it needs 10 to 15 minutes of warm-up before it meets the
specified performance.
For cleaning the instrument, use a soft and dry cloth.

11.1 Calibration Apparatus

Voltage/current generator (Yokogawa 7651 or the equiva-
lent): 1

A counter(Yokogawa TC100 or the equivalent) or oscillo-
scope (Yokogawa DL1540 or the equivalent): 1

A resistor (1k

, 1.6 k

): 1 each

6V battery: 1

11.2 Calibration Procedure

Connect the instruments as shown in Fig.11.1.  First adjust the out-
put-1 signal and then the output-2 signal.
Produce input signals equivalent to 0, 25, 50, 75 and 100% of the in-
put span to the converter from the voltage and current generator.
Then, check that the converter's output signal shows the outputs cor-
responding to 0, 25, 50, 75, and 100% of the input span within the
rated accuracy range. Take the same calibration procedure for the
output-2, and confirm that the frequency of the output 2 is the same
as that of the output 1.
When using a oscilloscope, the wave shaping of output pulse can be
confirmed.
If the output signal is out of the rated accuracy range, adjust the out-
put signal level using the Handy Terminal (JHT200).  For adjust-
ment, refer to Instruction Manual for VJ77 Parameters Setting Tools
(IM 77J01J77-01) or for Handy Terminal (IM JF82-01).

11

10

8

L+

N–

GND

5

2

1.6k

1k

+

9

7

1.6k

1k

+

10

11

3

2

1

4

5

6

7

8

9

Input signal

3

1

+

R

R:100

Externally connected for current input

Shunt resistor
(Externally connected for current input)

Voltage and

current

Generator

Calibration for output-1

Calibration for output-2

6V
Battery

6V
Battery

Counter or

Osilloscope

Counter or

Osilloscope

Power supply

Fig. 11.1