IM 01C25H01-01E
8-3
8. GENERAL SPECIFICATIONS
Load (Output signal code D and E)
0 to 1290
for operation
250 to 600
for digital communication
Communication Requirements “
”
(Safety approvals may affect electrical require-
ments.)
BRAIN
Communication Distance
Up to 2 km (1.25 miles) when using CEV polyethylene-
insulated PVC-sheathed cables. Communication distance
varies depending on type of cable used.
Load Capacitance
0.22
F or less
Load Inductance
3.3 mH or less
Input Impedance of communicating device
10 k
or more at 2.4 kHz.
HART
Communication Distance
Up to 1.5 km {1 mile} when using multiple twisted pair
cables. Communication distance varies depending on type
of cable used.
Use the following formula to determine cable length for
specific applications:
L= –
65
10
6
(R
C)
(C
f
+ 10,000)
C
Where:
L = length in meters or feet
R = resistance in
(including barrier resistance)
C = cable capacitance in pF/m or pF/ft
C
f
= maximum shunt capacitance of receiving devices in
pF/m or pF/ft
EMC Conformity Standards
,
EN 61326, AS/NZS CISPR11
PHYSICAL SPECIFICATIONS
Process connections
See the following table.
Table 8.2 Flange size and rating
Size
4-inch
3-inch
Flange
3-inch
2-inch
1
1
/
2
-inch
*1
JIS 10K, 20K, 40K, 63K
*2
ANSI Class 150, 300, 600
JPI Class 150, 300, 600
DIN PN10/16, 25/40, 64
Process
connection style
Flush type
High pressure
side: 4-inch
Low pressure
side: 3-inch
JIS 10K, 20K, 40K
*2
ANSI Class 150, 300
JPI Class 150, 300
DIN PN10/16, 25/40
JIS 10K, 20K
ANSI Class 150, 300
JPI Class 150, 300
DIN PN10/16, 25/40
Extended type
Combination type
(Extended and
Flush)
*3
T0804.EPS
*1: Flushing connection rings are always attached.
*2: Applicable for EJX438A only.
*3: Applicable for EJX118A only.
Gasket Contact Surface
See the following table.
Table 8.3 Gasket contact surface
T0805.EPS
SW,
SE,
SY
HW,
TW,
UW