<4. OPERATION>
4-3
IM 12T03A01-02E
4.2 Operation
4.2.1
Operation Mode
Operation of the Model DM8 vibration type density meter ruled by the mode as shown in Table
4.1. Measuring mode (DSPL), calibration mode (CAL) and maintenance mode (MAINT) are
transferred with the mode transfer switch.
Table 4.1 Operation Modes and Function No.
MODE SW FUNCTION
NO.
Contents
Display
Measuring
mode
(DSPL)
No display
1
2
Displays density (at reference temperature)
Displays density (measured value)
Displays measuring liquid temperature
g/cm
3
g/cm
3
°C
Calibration
mode (CAL)
3
4
5
6
7
8
9
A
Sets density of calibration solution 1
Sets temperature of calibration solution 1
Sets temperature coefficient of calibration solution 1
Starts calibration 1 (press SET key)
Sets density of calibration solution 2
Sets temperature of calibration solution 2
Sets temperature coefficient of calibration solution 2
Starts calibration 2 (press SET key)
□.□□□□
g/cm
3
□□□.□
°C
-□□□
x10
-5
g/cm
3
/°C
CAL-1
(displays density after calibration)
□.□□□□
g/cm
3
□□□.□
°C
-□□□
x10
-5
g/cm
3
/°C
CAL-2
(displays density after calibration)
Maintenance
mode
(MAINT)
1.
2.
3.
4.
5.
6.
7.
8.
9.
A.
B.
C.
D.
Cancellation of protection (enter 77)
Displays frequency (density)
Displays frequency (temperature)
Checks output signal
Checks LED (press SET key)
Sets low limit of the output range
Sets high limit of the output range
Sets reference temperature of measuring liquid
Sets temperature coefficient of measuring liquid
Sets detector constant A
Sets detector constant B
Displays detector calibration coefficient C (span)
Displays detector calibration coefficient D (zero)
7 7
Hz
Hz
□□□
%
- LED - (flash 5 times)
□.□□□□
g/cm
3
□.□□□□
g/cm
3
□□□.□
°C
-□□□
x10
-5
g/cm
3
/°C
□□□□□
□□□□□
(1) Measuring mode (DSPL): Mode for normal operation
An analog output in this mode is density value at the reference temperature.
Moreover, density value at measuring liquid temperature and temperature of measuring liquid are
digitally output. The density value at the reference temperature (without function No. display), the
density value at measuring liquid temperature (No.1) or the measuring liquid temperature (No.2)
can be displayed by selecting function No. Use FUNCTION key to select the FUNCTION No.
(2) Calibration mode (CAL): Mode for calibration. Sets or specifies data required
for calibration.
During this mode the analog and digital outputs are held (at the values just before the holding).
However, digital data output stops during calibration function No. ‘6’ or ‘A’.
When calibration computation ends holding is released and measured data are again output until
function No. ‘6’ or ‘A’ is transferred to another No.
5th Edition : Dec. 11, 2015-00