< 7. Calibration >
7-7
IM 12E01B30-02EN
1st Edition : Mar. 08, 2023-00
(3) Press on Main screen and go to the calibration menu.
(4) On the calibration menu, select Zero calibration.
(5) Confirm the reading is stable. Press [check stability].
(6) After the message pops saying the reading is stable, confirm that the calibration value is
zero. Then press [Next].
(7) Check the result of calibration and press [Accept] to update the Zero of the calibration
parameter.
(8) If Light source has been turned off, turn on the light source.
To shift to the slope calibration, proceed to the steps of “7.6 Slope calibration (Check tool)”
or “7.7 Slope calibration (Standard solution)”
. If you finish the operation, fix the clear cover
and feed the sample with the top cover on. Turn off the maintenance mode to end the
calibration.
7.5
For zero calibration by turning off the light
source
This can be selected only when the measured turbidity exceeds 200 NTU.
Refer to “7.4 Zero Calibration” for step.
7.6
Slope calibration (Check tool)
When you calibrate both zero point and slope, first calibrate the zero point and then the slope.
[How to perform Slope calibration-Check tool]
(1) Turn on the maintenance mode. Drain all the sample inside the Flow cell. Mount the check
tool referring to “7.2 Turbidity standards”.
(2) Press on Main screen and go to the calibration menu.
(3) On the calibration menu, select [Slope calibration (Check tool)] of turbidity manual calibration.
(4) Confirm the reading is stable. Press [check stability].
(5) After the message pops saying the reading is stable, confirm that a value of the check tool is
correctly displayed. Then press [Next].
(6) Check the result of calibration and press [Accept] to update the Slope of the calibration
parameter.
(7) Remove the check tool. If you finish the calibration, run the sample and turn off the
maintenance mode.
7.7
Slope calibration (Standard solution)
Normally slope calibration is performed with a check tool. If you perform with a standard solution,
follow the steps as below. If you do both Zero and Slope calibration, perform Zero calibration first
then do Span calibration. Prepare a standard solution referring to “
(1) Make 20 L or more of turbidity standard solution for slope calibration. (see 7.3 to 7.5)
Dilute the turbidity standard solution as necessary.
(2) Turn on the maintenance mode.
(3) Press on Main screen and go to the calibration menu.
(4) Select [Slope calibration (standard solution)] on the calibration menu.