<2. INSTALLATION, PIPING, AND WIRING>
2-15
IM 12B01B30-02EN
1st Edition : Mar. 08, 2023-00
CAUTION
Be sure to shut off the power supply prior to working.
(1) Loosen the four static screws and remove the detector Top window.
(2) Mount one each of the supplied desiccants on the top and sides of the top cover. Mount the
desiccant so that the printed side is facing the cover. Mount the top desiccant toward the
(3) Tighten the four screws to attach the top cover. The tightening torque of the screw is 1.5
N·m.
(4) This completes the replacement.
After mounting the desiccant, warm up the dryer for at least an hour to ensure that the desiccant
absorbs moisture in the detector before starting the measurement.
The side with no print is the moisture-absorbing surface.
Figure 2.17
Moisture absorption surface of the desiccant
Mount the desiccant on the top side to the right.
Figure 2.18
Mounting Desiccant to Top cover