3.6
Calibrating the Probe (Phase Correction)
Items Required
The following items are required.
Compensation Signal
• Frequency:
1 kHz
• Voltage (waveform amplitude):
1 Vp-p
• Waveform type:
Rectangular wave
• Output impedance:
1M
Ω
• Recommended signal:
CAL signal of DL1700 Series Digital Oscilloscope
(YOKOGAWA)
Waveform Monitor
• Frequency characteristics:
DC to 250 MHz (–3 dB point)
• Input coupling:
DC
• Input impedance:
50
Ω
• Recommended instrument:
DL1700 Series Digital Oscilloscope (YOKOGAWA)
The connection procedure and operation when the recommended signal is connected to
the recommended instruments are described below.
Connecting the instrument
CAUTION
• Do not apply a voltage exceeding the maximum input voltage to the signal input
terminal of the TA720. This may cause damage to the input section.
• Do not apply external voltage to the CAL signal output terminal of the DL1700
Series Digital Oscilloscope or the monitor output terminal of the TA720. This
may cause damage to the internal circuitry.
Check that the TA720 and the DL1700 Series Digital Oscilloscope are turned OFF and
connect them as shown in the figure.
1.
Connect the monitor output terminals on the rear panel of the TA720 (CH A and
CH B) to the signal input terminals of the DL1700 Series Digital Oscilloscope using
BNC cables.
For details on how to use the monitor output of the TA720, see section 13.5, “
Monitor Output.”
2.
Connect the BNC end of the probe to be phase corrected to the signal input
terminal of the TA720.
3.
Connect the other end of probe to the CAL signal output terminal of the DL1700
Series Digital Oscilloscope and the ground wire to the functional ground terminal.
DL1740
Monitor output
terminal
BNC cable
TA720
Signal input
terminal
Phase correction hole
CAL signal output terminal
Functional ground terminal