Repeat Auto Trigger
Taking the maximum and minimum values of the input signal during the first
measurement block to be 100% and 0%, respectively, the trigger level is set to the
specified percentage level. Unlike the single auto trigger, the trigger level is reset for
each measurement block.
Phase Adjustment for A-to-B Time Interval Measurements (Phase Adj) (see section 6.7
for the operating procedure)
The phase of CH B can be shifted with respect to CH A for the following measurement
functions: A-to-B time interval, period A & A-to-B time interval, pulse width A & A-to-B
time interval (pulse width A
→
A-to-B time interval). The phase shift can be used to
correct the offset of the phase of the CH B input signal with respect to the CH A input
signal due to causes such as the differences in the cable when measuring the A-to-B
time interval. The selectable range is 0.0 to 10.0 ns (resolution: 0.1 ns).
Reference Clock Selection (Ref. Clock) (see section 6.8 for the operating procedure)
The TA720 performs measurements using the internal 10-MHz reference clock as the
time reference. An external 10-MHz clock signal can be used in place of the internal
clock.