< 4. OPERATION >
4-3
IM 12A0V2-E
4.3 ST401G-PH6-
¨
-A, ST401G-PH7-
¨
-A
Adjust ball valve (BV1) and make the flow rate into pH flow-though type holder between 3 and 10
L/min.
Refer to Figure 4.3 for piping.
<Tubing materials>
X
X
Ø12 / Ø9 Polyethylene tube
(VP40)
Drain
VP16 pipe
pipe
VP40
PR1
: Pressure-reducing valve
BV1
: Ball valve
PHC
: pH converter
PHH
: Flow-through type holder
PHS
: pH sensor
TK1
: KCl tank
ULC
: Ultrasonic oscillator (Suffix code: -U)
X
(VP16)
Measuring water
PVC tray
Ø22 / Ø15 soft tube with net
PHC
TK1
UCL
PHS
PHH
BV1 PR1
Figure 4.3
Piping diagram of ST401G-PH6-
¨
-A, ST401G-PH7
¨
-A
(With pH converter, (ultrasonic oscillator))