CH1 to CH16 (LOGIC, EVENT, DSP)
Turn ON/OFF the channel
Section 5.1
Variable/Position
Sections 5.4 and 5.9
Turn Variable ON to set the upper and lower
limits of the vertical axis according to the
displayed waveform and zoom or expand the
waveform vertically. When Variable is turned
OFF, the vertical position of the waveform can
be specified.
Input coupling
Section 5.5
Select the method of coupling the input signal
to the vertical control circuit from AC and DC.
To check the ground level, select GND. To
measure rms values using the 701260, select
AC-RMS or DC-RMS. To measure the
temperature using the 701261, 701262, or
701265 (see the next page), select TC. Select
ACCL when measuring acceleration on the
701275 (see page 32).
Probe type
Section 5.6
Select the voltage probe attenuation or current
probe output rate according to the type of
probe used.
Bandwidth
Section 5.7
Select the high frequency band to be removed
from the input signal.
Vertical Zoom/Expand and Offset
Sections 5.8 and 5.10
Set the magnification and display range when
waveforms are zoomed/expanded vertically
and the offset voltage.
Voltage Modules
• High-Speed 10 MS/s, 12-Bit Isolation Module
MODEL: 701250, abbreviated name: HS10M12
• High-Speed High-Resolution 1 MS/s, 16-Bit
Isolation Module
MODEL: 701251, abbreviated name: HS10M16
• High-Speed 10 MS/s, 12-Bit Non-Isolation Module
MODEL: 701255, abbreviated name:
NONISO_10M12
• High-Voltage 100 kS/s, 16-Bit Isolation Module
(with RMS)
MODEL: 701260, abbreviated name: HV (with RMS)
• Universal (Voltage/Temp.) Module (Only When
Measuring Voltage)
MODEL: 701261, abbreviated name: UNIVERSAL
• Universal (Voltage/Temp.) Module (with AAF)
(Only When Measuring Voltage)
MODEL: 701262, abbreviated name: UNIVERSAL
(AAF)
• Temperature, High Precision Voltage Isolation
Module
(only when measuring voltage)
MODEL: 701265, abbreviated name: TEMP/HPV
• Acceleration/Voltage Module (with AAF)
MODEL: 701275, abbreviated name: ACCL/VOLT
Note
Zoom/Expand Vertically
The following two methods are available.
• Zooming in or out by setting the magnification
You can expand or reduce the waveform display
vertically by a factor in the range of
×
0.1 to
×
100.
The waveform display can be zoomed around the
vertical position.
• Zooming vertically according to the upper and
lower limits of the display range
You can zoom in on the desired section of the
observed waveform by specifying the upper and
lower limits of the vertical axis to change the display
range to a narrower range for each displayed
waveform. Conversely, you can widen the display
range to view waveforms outside the display range.
Note
Turn ON/OFF linear scaling
Section 5.11
Set the method of scaling the measured values.
Select OFF when not scaling.
Turn ON/OFF inverted waveform display
Section 5.12
Waveform label
Section 8.10
You can arbitrary set the waveform label using
up to eight characters.
to