9-3
IM DL350-01EN
Vertical Cursors (Vertical) - T-Y waveforms
Two straight dashed lines appear on the vertical axis (these are the vertical cursors). You can measure the time
between the trigger position and each cursor, the time difference between the two cursors, and the reciprocal
of the time difference between the two cursors. You can also measure the vertical signal value at each cursor
position and the level difference between the two cursors.
Cursor1
Cursor2
Δ
X
X1
X2
Measurement Source Waveform (Trace)
Set the measurement source waveform to one of the waveforms below.
ALL, CH1 to CH6
1
, 16chVOLT
2
, 16chTEMP/VOLT
2
, CAN
3
, LIN
3
, SENT
3
, Math1 to Math2
1 You can select the channel of an installed module. On a 4-CH module, select sub channel 1 or 2.
2 When a 16-CH voltage input module or 16-CH temperature/voltage input module is installed. After you select
16chVOLT or 16chTEMP/VOLT, select a sub channel.
3 On a model with the /VE option when a CAN bus monitor, CAN & LIN bus, or SENT monitor module is
installed. After you select CAN, LIN, or SENT, select a sub channel.
Moving the Cursors (Cursor1/Cursor2)
Use Cursor1 and Cursor2 to move the cursors.
You can set in the range of –5 to +5 div from the center of the waveform display window. The setting resolution
varies depending on the display resolution.
• Scope mode: Set the positions in divisions.
• Recorder mode: Set the positions in terms of the time (ddhhmmss) corresponding to divisions.
The start and end points are similar to the settable range of the cursor display position in cursor measurement.
For details, see “Selectable Range of Cursor Positions.”
Linking Cursors
You can move the cursors without changing the interval between Cursor1 and Cursor2.
Measurement Items (Item Setup)
You can measure the following horizontal values at the cursor positions.
X1
Time value at Cursor1
X2
Time value at Cursor2
ΔX
Difference between the time values of Cursor1 and Cursor2
1/ΔX
Reciprocal of the difference between the time values of Cursor1 and Cursor2
Y1
Vertical value at the intersection of Cursor1 and the waveform
1
Y2
Vertical value at the intersection of Cursor2 and the waveform
2
ΔY
Vertical value at the intersection of Cursor1/Cursor2 and the waveform
2
1 When Trace is set to ALL, the values for the channels of all installed modules, the sub channels, and the
Math channel are measured.
2 This option does not appear when Trace is set to ALL.
9 Cursor Measurement