2.  Measurement

 (refer to IM 12D03D02-01E,

Chapter 3)

Measuring liquid’s temperature should be between 0
and 80

8

C.

To fix the measuring range, see IM 12D03D02-01E, Sec.
4.3, Item (3).

 How to immerse the sensor.

 Measurement display

When the sensor is immersed into the measuring
liquid, the conductivity value is displayed. Each time

the 

MEAS

 key is pressed, the display toggles among the

three display types shown below.

3.  Maintenance

 Washing Sensor (refer to IM 12D03D02-01E,

Sec. 5.2)

After the measurement, rinse the sensor with clean
water (for example, tap water) to remove stains.

Calibration with Standard Solution

If sensor is not clean after long time use, use standard
solution and recalibrate the conductivity meter to
check if cell constant is normal.

 Replacing Batteries ( when 

  is flashing)

(Refer to IM 12D03D02-01E, Sec. 2.1)

When replacing batteries, replace both batteries at the
same time.
After replacing, reset the time.

 Replacing O-rings and gaskets (refer to IM

12D03D02-01E, Sec. 5.5)

If sensor connecter O-ring or battery box gasket
become dirty or damaged,
waterproofing is not assured. Then replace to new
ones.

How to store the measured value

 

 

HOLD (temporary storage)

 

 

How to store data in memory

To check the stored data, refer to IM 12D03D02-01E,
Sec. 4.3, Item (1).

4. Error Messages

 (refer to IM 12D03D02-01E,

Sec. 6.2)

Refer to IM 12D03D02-01E, Sec. 6.2 for corrective
action.

5. Storage

 (refer to IM 12D03D02-01E, Sec. 5.5)

· Wash sensor with water before storing.
· Leave sensor connected to meter to keep connector

clean.

· Remove batteries from meter if it is to be stored for long

time.

Air vent

Measuring liquid level

Cover

If air bubbles are on this section, 
measurement error may occur.

Electrode 
element

F04.EPS

F05.EPS

Standard display

Calendar display

Clock display

MEAS

MEAS

MEAS

day month year

Measuring liquid 
temperature

Temperature 
compensation coefficient

This display appears when NaCl 
solution is specified at temperature 
compensation coefficient setting.

conductivity

hour : minute

Delete stored data

     Press
              to 

delete stored data.

Display day and 
month of stored 
data 

Display 
dAt with
     or
key.

Press 

key to 
set.

Press 

key to abort 

data deletion. 

Press          key to return to 

measurement mode.

Display 
other data.

Display year of stored data

Display time of stored data

F/ENT

F/ENT

F/ENT

F/ENT

F/ENT

DATA

DATA

DATA

DATA

DATA

DATA

Stored measurement data

Latest 
data

MEAS

F08.EPS

F06.EPS

or

HOLD

HOLD

MEAS

F07.EPS

Abort data storage.

Execute data storage.

Data is not stored when        key is not pressed.

Press          key to store data while           is flashing.

While          is flashing, press
or         keys to abort data storage.

F/ENT

F/ENT

MEAS

DATA

DATA

T03.EPS

Error message

Err1

Err2

Err3

Err6

or

                            Name

Temperature compensation computation value out of range

Out of temperature measuring range

Calibration error

Meter electronics failure

Out of measuring range