IM 12B03D02-01E
5-10
5
.
Keypad and Display Functions
(9) Set beep on/off (bZ.o) panel
The beep sound on key press can be enabled/disabled in this panel. Use
keys
to select on or off and press
F/ENT
key to confirm. Note that this beep setting does
not affect the alarm sounding (see Item (7)).
F050309.ai
F/ENT
to turn ON (enable).
F/ENT
to turn OFF (disable).
F/ENT
(10) Standard solution setting (Std) panel
Used to select the type of standard solutions: NIST or US. The factory default is
NIST. When using standard solutions prepared in accordance with the Japanese
standards, use the meter with the factory default of NIST. Select US only when using
standard solutions prepared in accordance with the U.S. standards (refer to Chapter
4, “Calibration”).
F050310.ai
to toggle
F/ENT
to confirm
F/ENT
(11) Initialize calibration parameters (I.CP) panel
Used to initialize the parameters saved by calibration to default settings: slope at
1.000 and asymmetry potential at 0.0 mV.
F050311.ai
to toggle
to
proceed
F/ENT
to cancel
F/ENT
to execute
F/ENT