8. Technical Information
8.2 Relationship between EMF of Glass Membrane and pH
Value
The relationship between the potential difference (electromotive force) developed across
the glass membrane and the pH value had been studied and the theoretical values were
determined. Actual values, however, do not match the theoretical values due to
manufacturing variations or deterioration with time. Therefore, a pH meter must be
calibrated using standard solutions.
The emf of a glass electrode is affected by temperature. To compensate for this
temperature effect is called “temperature compensation” which is essential for pH
measurement.
Figure 8.2 shows the diagram of glass electrode membrane. Both membrane surfaces in
contact with solutions are hydrated and the hydrogen ion activity in these hydrated layers
is constant. A boundary potential develops depending on the ratio of the hydrogen ion
activities of the hydrated layer and of the solution. This boundary potential, e, is
expressed from the Nernst equation as follows.
R : gas constant, 8.3145 [J/(mol·K)]
T : absolute temperature (t [
8
C]
1
273.15) [K]
F :
Faraday constant, 9.6485
3
10 [C/mol]
C :
potential at interface between glass and internal solution
C :
potential at interface between glass and sample solution
e
5
2
Where:
pH
1
C (internal solution side)
2.3026 R T
F
(8.1)
i
i
i
e
5
2
pH
1
C (sample solution side)
2.3026 R T
F
(8.2)
S
4
i
S
S
S
Given the potential at internal solution side is reference, the difference across the
membrane e
g
is:
e
5
e
2
e
5
(pH
2
pH )
1
(C
2
C )
2.3026 R T
F
(8.3)
g
i
i
i
S
S
S
To determine the difference of membrane potentials, two inner electrodes are
incorporated in a glass electrode and a reference electrode and the difference in potential
at two electrodes is measured by a pH converter with high input impedance. This
potential difference Eg is expressed as follows when the difference of single electrode
potential of the two inner electrodes and C
S
– C
i
in equation 8.3 are collectively
represented by E
AS
.
E
5
(pH
2
pH )
1
E
2.3026 R T
F
(8.4)
g
i
S
AS
E
5
(54.20
1
0.1984 t)
3
(pH
2
pH )
1
E
(8.5)
g
i
S
AS
In the equation, pH
i
is the pH buffer solution filled in the glass electrode so the pH
should be constant. Using a solution (pH standard solution) with known pH as pH
S
at a
certain temperature, the relationship between millivolt and pH can be determined,
thereby the pH can be directly derived from the membrane potential difference, as shown
in Figure 8.3.
Summary of Contents for PH71
Page 1: ...User s Manual IM 12B03D01 01E Model PH71 Personal pH Meter IM 12B03D01 01E 3rd Edition ...
Page 7: ...IM 12B03D01 01E 6 Contents ...
Page 17: ...IM 12B03D01 01E 1 10 1 Outline ...
Page 23: ...IM 12B03D01 01E 2 6 2 Preparation ...
Page 35: ...IM 12B03D01 01E 4 8 4 Calibration ...
Page 49: ...IM 12B03D01 01E 5 14 5 Keypad and Display Functions ...
Page 57: ...IM 12B03D01 01E 6 8 6 Maintenance ...
Page 73: ...IM 12B03D01 01E 8 10 8 Technical Information ...
Page 79: ...IM 12B03D01 01E 6 Appendix ...
Page 81: ...IM 12B03D01 01E MSDS 2 MSDS ...
Page 82: ...IM 12B03D01 01E MSDS 3 MSDS ...
Page 83: ...IM 12B03D01 01E MSDS 4 MSDS ...
Page 84: ...IM 12B03D01 01E MSDS 5 MSDS ...
Page 85: ...IM 12B03D01 01E MSDS 6 MSDS ...
Page 86: ...IM 12B03D01 01E MSDS 7 MSDS ...
Page 87: ...IM 12B03D01 01E MSDS 8 MSDS ...
Page 88: ...IM 12B03D01 01E MSDS 9 MSDS ...
Page 89: ...IM 12B03D01 01E MSDS 10 MSDS ...
Page 90: ...IM 12B03D01 01E MSDS 11 MSDS ...
Page 91: ...IM 12B03D01 01E MSDS 12 MSDS ...
Page 92: ...IM 12B03D01 01E MSDS 13 MSDS ...
Page 93: ...IM 12B03D01 01E MSDS 14 MSDS ...
Page 94: ...IM 12B03D01 01E MSDS 15 MSDS ...
Page 95: ...IM 12B03D01 01E MSDS 16 MSDS ...