HART Communication
30
TI 12A01A01-60EN
SC450G
Parameter
R/W
Parameter
Value
Unit
Note
Func
[Type: Control]
R/W
P
PI
PID
Param
[Type: Control]
R/W
NONE
CONDUCT
[Measure: Cond only,
Cond + Conc]
RESIST1 [Measure:
res only]
CONCENT1
[Measure: Conc only]
TEMPERATURE
CONDUCT2
[Measure: Cond only]
RESIST2 [Measure:
res only]
CONCENT2
[Measure: Conc only]
CONCENT [Measure
:Cond + Conc]
PID SP
[Type: Control]
R/W
Numeric value
no limit
S/cm, S/m [Param:
CONDUCT1, 2]
p p b [ P a r a m :
CONCENT1]
O.cm, O.m [Param:
RESIST1, 2]
ッ
-C,
ッ
-F [Param:
TEMPERATURE]
% [ P a r a m :
CONCENT2]
PID Rng
[Type: Control]
R/W
Numeric value
greater than
0.000000
S/cm, S/m [Param:
CONDUCT1, 2]
p p b [ P a r a m :
CONCENT1]
O.cm, O.m [Param:
RESIST1, 2]
ッ
-C,
ッ
-F [Param:
TEMPERATURE]
% [ P a r a m :
CONCENT2]
PID Dir
[Type: Control]
R/W
direct
reverse
PID MR
[Type: Control (Func: P)]
R/W
Numeric value
0.000000 to
100.000000
%
PID I time
[Type: Control (Func: PI, PID) ]
R/W
Numeric value
1.000000 to
3600.000000
s
PID D time
[Type: Control (Func: PID)]
R/W
Numeric value
0.000000 to
60.000000
s
Analog output
[Type: Control]
R/W
PF
DC
Max freq
[Type: Control (Analog output: PF)]
R/W
Numeric value
1.000000 to
120.000000
p/m
DC time
[Type: Control (Analog output:
DC)]
R/W
Numeric value
1.000000 to
1800.000000
s
Expire time
[Type: Control]
R/W
Numeric value
1.000000 to
1800.000000
s
Func
[Type: Alram]
R
Alarm
Oct. 03, 2017-00