IM 701974-01E 2/2
4 Specifications
Item
Specifications
Standard Operating
Environment
Temperature: 23°C ±5°C
Humidity:
55% ±10% RH (no condensation)
System Requirements
Temperature: 5°C to 40°C
Humidity:
20% to 80% RH (no condensation)
Storage Environment
Temperature: –20°C to 60°C
Humidity:
20% to 80% RH (no condensation)
Total length of probe
Approximately 1.1 m
Connector type
SMA
Input resistor
1
Within ±2% of 450 Ω
Within ±2% of 950 Ω
Input capacitance
450 Ω: 0.25 pF (Typical
2
)
950 Ω: 0.4 pF (Typical
2
)
Attenuation ratio
3
Within ±3.5% of 10:1
Within ±3.5% of 20:1
Frequency bandwidth
4
DC to 5 GHz (–3 dB or more)
Rise time
4
70 ps or less (Typical
2
)
Maximum input voltage
20 Vrms, 40 VACpeak
Nondestructive input voltage
20 Vrms, 40 VACpeak
1 When no load is present. Standard operating environment.
2 Typical value represents a typical or average value. It is not strictly guaranteed.
3 In combination with an oscilloscope having an input resistance of 50 Ω. Standard operating environment.
4 When the distance between probe tip and ground is 9 mm. Standard operating environment.