YOKOGAWA PBL5000 User Manual Download Page 2

IM 701974-01E  2/2

4 Specifications

Item

Specifications

Standard Operating 

Environment

Temperature:  23°C ±5°C

Humidity:  

55% ±10% RH (no condensation)

System Requirements

Temperature:  5°C to 40°C

Humidity:  

20% to 80% RH (no condensation)

Storage Environment

Temperature:  –20°C to 60°C

Humidity:  

20% to 80% RH (no condensation)

Total length of probe 

Approximately 1.1 m

Connector type 

SMA

Input resistor

1

Within ±2% of 450 Ω

Within ±2% of 950 Ω

Input capacitance

450 Ω: 0.25 pF (Typical

2

)

950 Ω: 0.4 pF (Typical

2

)

Attenuation ratio

3

Within ±3.5% of 10:1

Within ±3.5% of 20:1

Frequency bandwidth

4

DC to 5 GHz (–3 dB or more)

Rise time

4

70 ps or less (Typical

2

)

Maximum input voltage

20 Vrms, 40 VACpeak

Nondestructive input voltage

20 Vrms, 40 VACpeak

1  When no load is present. Standard operating environment.

2  Typical value represents a typical or average value. It is not strictly guaranteed.

3  In combination with an oscilloscope having an input resistance of 50 Ω. Standard operating environment.

4  When the distance between probe tip and ground is 9 mm. Standard operating environment.

Reviews: