background image

IM MY40-02EN <P 2>

  <Deleting data>

Data saved can be collectively deleted.

(1)  Turn the function switch from power-off to the continuity test position (Clr) while 

 

pressing and holding the 

 key (MEM).

(2)  “MEM” and “CLr” appear.

(3)  Press the 

 key (MEM). “CLr” starts flashing.

(4)  Press the 

 key (MEM) again while “CLr” is flashing. 

 

The buzzer sounds and all data are deleted (do not change the position of

 

the function dial for two seconds when “CLr” is displayed). 

 

The tester changes its mode to measurement of conductor resistances.

NOTE

 

To stop deleting of data:

Do not perform any key operation and wait for 10 seconds while  

the “CLr” indication mentioned in step (3) is flashing; or 

 

Turn the function switch to another rating.

  <Turning off the memory feature>

Turn the function switch to the off position.

14.8  Comparator

When a measured insulation resistance is less than the reference value setting,  

the 

 mark appears and the buzzer sounds. Also the measured value is 

 

automatically held for 5 seconds.

  <Selecting the reference value>

Select from three preset values for each rating.

Default values: no.01: 0.1 MΩ/ no.02: 0.2 MΩ/ no.03: 0.4 MΩ

(1)  Turn the function switch to a measurement rating.

(2)  Press the 

 key.

(3)  “ 

 ”appears and the reference value appears on the subdisplay.

(4)  Pressing the 

 key (Select) changes the reference value display as follows.

 

 

0.1 MΩ → 0.2 MΩ → 0.4 MΩ → Comparator off (no display)

 

The reference value currently displayed is selected as the comparator setting.

  <Setting the preset values>

Main display

Sub-display

Position of decimal point

Measured

value

Reference

value

Reference

value

no.

Maine

display

Sub-

display

Comparator setting

Insulation resistance 

measurement 

(with comparator)

  <Setting the preset values>

Each of the default preset values no.01 to no.03 can be changed independently.

(1)  Turn the function switch from power-off to a desired rating while pressing and 

 

holding the 

 key. “no.01” on the sub-display flashes. 

 

Press the 

 key (Select) to select no.02 or no.03.

(2)  Confirm the setting no. with the 

 key (Enter).

 

The reference value on the main display can be set.

(3)  Select the position of the decimal point with the 

 key (Select).

(4)  Confirm the decimal point with the 

 key (Enter).

(5)  Select the number for each digit sequentially. 

 

Pressing the 

 key (Select) changes the number as follows. 

 

0 → 1 → 2 ··· 8 → 9 → 0 

(6)  Confirm the number with the 

 key (Enter).

(7)  Set the number for the next digit.

(8)  When settings for all the digits are complete, 

 

the reference value starts flashing.

(9)  Confirm the reference value with the 

 key (Enter). 

 

The buzzer sounds and the setting ends. 

 

The tester changes its mode to measurement of insulation resistances.

Repeat steps (1) to (9) to change other preset values.

NOTE

 

• To cancel the preset setting

Pressing the 

 key (Select) in step (8) returns the setting procedure to step (1). 

To stop the setting, turn the function switch to the off position.

• To set the preset values for another rating

Turn the function dial to the desired rating (insulation resistance) in step (8).

15.  Corrective Measures in Case of Failure

If the instrument does not operate properly after taking the corrective measures shown  

below or any other failure that is not covered here occurs, contact the vendor from which  

you purchased the instrument.

Failures

• The tester displays nothing after turning the function switch to the on position.

• The display is light in color.

• The measured values are erroneous.

Items to Check

• The batteries are installed properly and their power levels are not low.

• The measuring probes are connected correctly and are not damaged.

• Whether the failure reoccurs after turning off the power and retrying the operation.

Error Messages

Err.0

An internal operation failure.

Turn off the power and retry the operation.

* If the failure reoccurs, the tester needs repairing.

Err.1

Cannot save the comparator or memory settings.

Measurement functions other than the comparator and  

memory functions operate normally.

Settings and data saved without the error indication (Err.1) can be used.

* If the failure reoccurs, the tester needs repairing.

Err.2

An internal operation failure.

Turn off the power and retry the operation.

* If the failure reoccurs, the tester needs repairing.

13.  Measuring Conductor Resistance (Continuity)

The tester can measure the conductor resistance of 0 to 400 Ω. 

The buzzer beeps for resistances approximately 40 Ω or lower.

WARNING

 

Turn off the power to the measured object before connecting or 

 

measuring conductor resistance.

13.1  Setting the Function Switch

Turn the function switch to a conductor resistance measurement ( 

 ) position.

13.2  Connecting the Earth Probe

Securely connect the earth probe to the measured object.

13.3  Connecting the Line Probe

Bring the line probe into contact with the measured object. 

The display indicates the resistance of the measured object.

NOTE

 

You do not have to press the MEAS key when you measure conductor resistances.

14.  Additional Features

14.1  Double-action 1000-V Function

The tester is equipped with the RELEASE button to protect the measured circuit from 

 

damage due to inadvertent measurement at 1000 V.

To select the 1000 V rating, turn the function switch to the position  

while pressing this button. 

To cancel the 1000 V rating, turn the function switch to the off position or  

another rating position.

14.2  Live-line Alarm

Always turn off the power to the measured object before connecting or measuring 

 

insulation resistance. If an AC voltage of more than 40 V is applied,  

the ALARM LED flashes and the buzzer beeps. 

 

In this case, stop the measurement immediately and check the power supply voltage.

14.3  Locking the MEAS Key (for Continuous Measurement)

The MEAS key, when pulled up to the right, can be locked to ensure the key remains 

 

turned on. Use this mechanism when making continuous measurement over  

a prolonged period.  

Note, however, that leaving the key turned on for an unreasonably long time will  

accelerate the discharge of the batteries.

14.4  HOLD Feature

Measured insulation resistances are automatically held for approximately  

five seconds. Turning off the MEAS key initiates holding.

14.5  Auto Power-off

The tester is automatically turned off when no key operations are performed  

for 10 minutes.

The buzzer sounds 9 and a half minutes after the last key operation (at 1-sec intervals). 

 

If any measurement or key operation is not made while the buzzer sounds, 

 

the tester is turned off. However, it is not turned off when measurement is in  

progress or an alarm occurs.
To use the tester after the auto power-off is triggered,  

press the 

 key or the 

 key, or turn the function switch to the off  

position once before proceeding with the desired operation.

14.6  LIGHT and COMP Keys

The 

 key is used as the Select key for settings and storage numbers of 

 

the comparator and memory functions.

The 

 key is used as the Enter key for settings and storage numbers of 

 

the comparator and memory functions. 

Pressing the Enter key while the display is flashing, confirms selection of 

 

the setting or the storage numbers.

  <When setting the memory or comparator function>

Turn the function switch to an insulation resistance rating position while  

pressing and holding the 

 key or the 

 key until all the displayed  

digits stop flashing and remain on, and the buzzer sounds.

Or position the function switch to a rating of the insulation resistance first, 

 

press the 

 key or the 

 key when all the digits appear, 

and hold the key until the buzzer sounds.

  <Backlight>

• The backlight remains off while the comparator is set.

• The backlight is lit when memory is in use.

14.7  Memory Feature (Data Saving)

Up to 20 measured values of the insulation resistance for each rating can be  

saved to memory.

  <Displaying data>

(1)  Turn the function switch from power-off to a measurement rating 

while  

pressing and holding the 

 key (MEM).

 

“MEM” appears and “no.01” (storage number) flashes on the sub-display.

(2)  Select a storage number.

 

Press the 

 key (Select) to select a storage number (no.01 to no.20).

If an insulation resistance data is saved with the selected (displayed) number, 

 

the insulation resistance value is displayed.

If any data is not saved with the selected (displayed) number, “----” is displayed.

  <Saving data>

First select a storage number as in steps (1) and (2) above. 

Overwriting data is possible.

(3)  Perform measurement with the 

 key.

(4)  Turn off the 

 key.  

The measured value is held and starts flashing. 

 

Press the 

 key (Enter) while the measured value is flashing to save 

 

it to memory (it is held for five seconds in which time the next data cannot be saved).

NOTE

 

If the data being held is invalid, “----” is displayed and it cannot be saved. 
In this case, perform measurement again and save the data.

Reviews: