7-27
IM 04P01B01-01E
Setup Operations for Changing/Adding Functions (Basic Setting Mode)
1
2
3
4
5
6
7
8
9
10
11
12
App
Index
7.17 Initializing the Settings
This section explains the details of initializing the recorder settings to their factory default.
Be careful, because all settings except the date/time and the adjustment values of the
pen/dot printing position will be initialized.
Basic Setting mode cannot be entered when recording is in progress or when
computation is in progress on models with the computation function (/M1 option).
Procedure
1.
Hold down the
MENU
key for 3 seconds to enter Setting mode.
2.
Hold down the
and
keys simultaneously for 3 seconds to enter Basic
Setting mode.
3.
Press the
key or
SHIFT
+
key to select
Initialize
and then press the
key.
4.
Press the
key to select the settings to be initialized and then press the
key.
Set:
Initializes the settings in Setting mode.
Setup+Set: Initializes the settings in Basic Setting mode and Setting mode.
5.
To execute the initialization, press the
key to select
Yes
and then press the
key.
The settings are initialized, and the recorder returns to Operation mode.
To cancel the initialization, press the
ESC
key The screen returns to the
Basic=Initialize
screen.
If you press the
key to select
No
and press then the key, nothing hap-
pens.
Select the range to be initialized.
Select Yes and press
to execute
initialization and return to Operation
mode.
ESC/?
Basic=Initialize
Mode=Setup+Set
Are you sure?=Yes
Explanation
Items That Are Not Initialized
The following items are not initialized.
• Adjustment value of the pen position and the adjustment value of the printer carriage
position (Basic Setting mode)
• Date/Time (Setting mode)
Summary of Contents for mR10000 436101