background image

3

IM 77J01A04-01E

4th Edition

4.  MAINTENANCE

The product starts running immediately when the power is turned 

on; however, it needs 10 to 15 minutes of warm-up before it 

meets the specified performance.

4.1  Calibration Apparatus

 • A DC voltage/current standard (YOKOGAWA GS200 or the 

equivalent)

 •

A digital multimater (YOKOGAWA DM7561 or the equivalent)

 •

A precision resistor of 250 Ω ± 0.01%, 1 W

4.2  Calibration Procedure

1. 

Connect the instruments as shown below. First check the 

channel 1 signal and then the channel 2 signal.

  

R: 250 

 precision resistor

     for current output

R: 250 

 precision resistor

     for current output

Power supply

With SINK current

With SINK current

Without SINK current

Without SINK current

Output

Output

DC voltge/

current

standard

DC voltge/

current

standard

DC voltge/

current

standard

DC voltge/

current

standard

10

11

3 2 1

4

5

6

7

8

9

11

10

8

L

+

N

GND

3

9

+

Input

Channel 1

Channel 2

DMM

7

9

R

+

1

3

+

6

5

+

Input

4

6

+

DMM

2

5

R

+

2.  Use the DC voltage/current standard and apply input signals 

equivalent to 0, 25, 50, 75, and 100% of the input span to the 

product. 

3.  Check to see the corresponding output voltages are 0, 25, 50, 

75, and 100% respectively and within the specified accuracy 

rating. “R” is used for current output.

5.  SAFETY AND EMC STANDARDS

Safety:

IEC/EN 61010-1 compliance (CE)

IEC/EN 61010-2-030 compliance (CE)

CAN/CSA-C22.2 No.61010-1 (CSA)

CAN/CSA-C22.2 No.61010-2-030 (CSA)

UL61010-1 (CSA NRTL/C)

UL61010-2-030 (CSA NRTL/C)

Overvoltage category: II

Pollution degree: 2

Measurement category: O (other)

Rated power supply voltage:  15-30 V DC (±10%) 2.2 W or 

 

100-240V AC/DC (±10%) 

50/60 Hz 2.6 W  5.3 VA

Rated voltage (Input signal) : 25.5 V DC

Rated voltage (Analog output signal):  1 to 5 V DC, 

 

4 to 20 mA DC

Rated transient overvoltage: 1500 V (*)

* This is a reference safety standard value for the 

measurement Category O. This value is not necessarily 

a guarantee of instrument performance.

 CAUTION

This instrument is for Measurement Category O 

(other). Do not use it for measurements in locations 

falling under Measurement Categories II, III, and IV.

    

Internal Wiring

Outlet

Entrance 

Cable

III

T

O

II

IV

Measurement 

category

Description

Remarks

O

(other) For measurements 

performed on circuits not 

directly connected to MAINS.

II

CAT.II For measurements 

performed on circuits directly 

connected to the low-voltage 

installation.

Appliances, 

portable 

equipments, 

etc.

III

CAT.III For measurements 

performed in the building 

installation.

Distribution 

board, circuit 

breaker, etc.

IV

CAT.IV For measurements 

performed at the source of 

the low-voltage installation.

Overhead 

wire, cable 

systems, etc.

EMC standards:

CE:

EN 61326-1 Class A, Table 2

EN 61326-2-3

* The instrument continues to operate at a measurement 

accuracy of within ±20% of the range during testing.

EN 55011 Class A, Group 1

EN 61000-3-2

EN 61000-3-3

RCM: 

EN 55011 Class A, Group 1

KC:

Electromagnetic wave interference prevention standard, 

electromagnetic wave protection standard

 CAUTION

Caution to comply with EMC standards: When 

operating this instrument by external power supply, 

use an independent power unit conforming to CE-

marking. Be sure to use the lightning arrester to 

comply EMC standards.

Note: When option code /C0 or /FB is specified, the conformity to 

the safety and EMC standards is excluded.