The product is sensitive to static electricity; 

exercise care in operating it. Before you operate 

the product, touch a nearby metal part to discharge 

static electricity.

4.  PART NAMES OF FRONT PANEL

The figure below shows the MA7D with its front panel (cover) 

being open.

  

Zero-adjustment volume
Span-adjustment volume

Power indicator lamp
Turns on at power on.

5.  MAINTENANCE

The product starts running immediately when the power is turned 

on; however, it needs 10 to 15 minutes of warm-up before it 

meets the specified performance. 

5.1  Calibration Apparatus 

 • A DC voltage/current standard (Yokogawa GS200 or the 

equivalent)

 • A digital multimater (Yokogawa DM7560 or the equivalent)

 • A precision resistor of 250 Ω ± 0.01%, 1 W 

5.2  Calibration Procedure 

(1) Connect the instruments as shown below.

(2) Use the DC voltage/current standard and apply input signals 

equivalent to 0, 25, 50, 75, and 100% of the input span to the 

product. Check to see the corresponding output voltages are 

0, 25, 50, 75, and 100% respectively and within the specified 

accuracy rating. “R” is used for current output. 

 • If the output signals are out of the accuracy rating range, 

adjust the output signal level using the zero and span 

adjustment volumes on front face of the product.

  

6 5

4 3

7 8

1 2

6

7

8

1

2

Ro

Ri

3

4

Power supply

Input

Output

Receiving resistor

for current input

Ro: 250 

Ω 

precision resistor

        for current output

Ri: External receiving resistor for

      current input (See chapter 3.)

2

1

2

1

L1

N2

GND

DC voltge/

current

standard

Digital

multimater

6.  TRANSPORT AND STORAGE CONDITIONS

 • Temperature: -25 to 70°C

 • Temperature change rate: 20°C per hour or less

 • Humidity: 5 to 95%RH (no condensation)

 CAUTION

Keep this product in a conductive bag when plugged 

out, during transport or storage.

3

IM 77J04H01-01E 3rd Edition  June 29, 2018-00