The product is sensitive to static electricity;
exercise care in operating it. Before you operate
the product, touch a nearby metal part to discharge
static electricity.
4. PART NAMES OF FRONT PANEL
The figure below shows the MA7D with its front panel (cover)
being open.
Zero-adjustment volume
Span-adjustment volume
Power indicator lamp
Turns on at power on.
5. MAINTENANCE
The product starts running immediately when the power is turned
on; however, it needs 10 to 15 minutes of warm-up before it
meets the specified performance.
5.1 Calibration Apparatus
• A DC voltage/current standard (Yokogawa GS200 or the
equivalent)
• A digital multimater (Yokogawa DM7560 or the equivalent)
• A precision resistor of 250 Ω ± 0.01%, 1 W
5.2 Calibration Procedure
(1) Connect the instruments as shown below.
(2) Use the DC voltage/current standard and apply input signals
equivalent to 0, 25, 50, 75, and 100% of the input span to the
product. Check to see the corresponding output voltages are
0, 25, 50, 75, and 100% respectively and within the specified
accuracy rating. “R” is used for current output.
• If the output signals are out of the accuracy rating range,
adjust the output signal level using the zero and span
adjustment volumes on front face of the product.
6 5
4 3
7 8
1 2
6
7
8
1
2
Ro
Ri
3
4
Power supply
Input
Output
Receiving resistor
for current input
Ro: 250
Ω
precision resistor
for current output
Ri: External receiving resistor for
current input (See chapter 3.)
2
1
2
1
L1
N2
GND
DC voltge/
current
standard
Digital
multimater
6. TRANSPORT AND STORAGE CONDITIONS
• Temperature: -25 to 70°C
• Temperature change rate: 20°C per hour or less
• Humidity: 5 to 95%RH (no condensation)
CAUTION
Keep this product in a conductive bag when plugged
out, during transport or storage.
3
IM 77J04H01-01E 3rd Edition June 29, 2018-00