background image

 CAUTION

 

The power line and input/output signal lines should 

be installed away from noise-generating sources. 

Other wise accuracy cannot be guaranteed.

 

Make sure to earth ground the ground terminal 

through minimum resistance. The length and 

thickness of the grounding cable should be as 

short and thick as possible.  Directly connect the 

lead from the ground terminal (terminal no. 6) of 

the product to the ground.  Do not carry out daisy-

chained inter-ground terminal wiring.

 

The product is sensitive to static electricity; 

exercise care in operating it. Before you operate 

the product, touch a nearby metal part to discharge 

static electricity.

4.  PART NAMES OF FRONT PANEL

The figure below shows the MB1 with its front panel (cover) being 

open. 

Zero-adjustment volume
Span-adjustment volume

Power indicator lamp
Turns on at power on.

5.  MAINTENANCE

The product starts running immediately when the power is turned 

on; however, it needs 10 to 15 minutes of warm-up before it 

meets the specified performance. 

5.1  Calibration Apparatus 

 •

A AC voltage current standard (YOKOGAWA 2558A or the 

equivalent)

 • A digital multimater (YOKOGAWA DM7560 or the equivalent)

 •

A precision resistor of 250 Ω ± 0.01%, 1 W

5.2  Calibration Procedure 

(1) Connect the instruments as shown below.

(2) Use the AC voltage current standard and apply input signals 

equivalent to 0, 25, 50, 75, and 100% of the input span to the 

product. Check to see the corresponding output voltages are 

0, 25, 50, 75, and 100% respectively and within the specified 

accuracy rating. “R” is used for current output. 

 • If the output signals are out of the accuracy rating range, 

adjust the output signal level using the zero and span 

adjustment volumes on front face of the product.

6 5

4 3

7 8

1 2

L+

N–

GND

6

7

8

1

2

R

+

3

4 ±

A

Power supply

Input

AC voltge 

current

standard

R: 250 

 precision resistor

for current output

DMM

Output

CT Protector

6.  TRANSPORT AND STORAGE CONDITIONS

 •

Temperature: −25 to 70°C

 •

Temperature change rate: 20°C per hour or less

 •

Humidity: 5 to 95%RH (no condensation)

 CAUTION

Keep this product in a conductive bag when plugged 

out, during transport or storage.

3

IM 77J04B01-01E

1st Edition