27
IM 12D06D05-01E
5
MENU
STRUCTURE
COMMISSIONING
Menu
Parameter Default
Range
values
min.
max.
Alarm S1
Setpoint
900.0 mS/cm (high)
- inf
+ inf
Alarm S1
Hysteresis
9.000 mS/cm
0.0 µS/cm
+ inf
Alarm S1
Delay Time
0.2 sec.
0.0 sec.
+ inf
Alarm S1
Expire Time
0.0 sec.
0.0 sec.
1800 sec.
Alarm S2
Setpoint
100.0 mS/cm (low)
- inf
+ inf
Alarm S2
Hysteresis
1.000 mS/cm
0.0 µS/cm
+ inf
Alarm S2
Delay Time
0.2 sec.
0.0 sec.
+ inf
Alarm S2
Expire Time
0.0 sec.
0.0 sec.
1800 sec.
S1 (control)
Expire Time
0.0 sec.
0.0 sec.
1800 sec.
P-control S1
Setpoint
500.0 mS/cm
- inf
+ inf
P-control S1
Range
100.0 mS/cm
- inf
+ inf
P-control S1
Manual Reset
0.000%
0%
100%
PID-control S1
I-time
3600 sec.
1.0 sec.
3600 sec.
PID-control S1
D-time
0.0 sec.
0.0 sec.
60 sec.
S2, S3, S4 Similar structure to S1