<6. SETTING AND CALIBRATION>
6-20
IM 11G02N01-01E
Description of setting items
To close “Setting of Auto Zero calibration”
6.5 Setting of auto zero calibration
6.5.1 Auto zero calibration
Auto zero calibration is automatically carried out at the time when zero calibration is set.
Components for which a calibration is to be made are determined by setting of auto calibration
component in Item 6.2.4.
Before changing the setting of auto zero calibration, set the ON/OFF to OFF.
MODE
ENT
( )
ENT
End of Auto Zero Calibration Setting
ENT
( )
F6_5_1.ai
(1) During measurement, press the
MODE
key
to display the User Mode.
(2) Point the cursor to “Setting of Auto Zero
Calibration” by pressing the
or
key. Press the
ENT
key.
(3) In the “Setting of Auto Zero Calibration”
screen that appears, point the cursor to
any item you want to set by pressing the
or
key. Press the
ENT
key.
(4) In the “Auto Zero Calibration Parameter
Setting” screen that appears, perform the
value entry or the setting. For the value
entry or setting change, use the
or
key. To change the setting, use the
key to move the cursor to the right.
After setting, press the
ENT
key, and auto
zero calibration is carried out by the
entered setting value.
• Start Time : Setting at the first calibration
(day of the week, hour, minute)
• Cycle :
A period between the start time of
one calibration and another
(unit : hour/day)
• Flow Time : The time required for the
calibration gas to be replaced in
the cell
• ON/OFF :
ON/OFF of auto zero calibration
To close the “Setting of Auto Zero calibration” or
cancel this mode midway, press the
ESC
key.
A previous screen will return.
Summary of Contents for IR400 series
Page 1: ...User s Manual R Model IR400 Infrared Gas Analyzer IM 11G02N01 01E IM 11G02N01 01E 5th Edition ...
Page 7: ...Blank Page ...
Page 11: ...Blank Page ...
Page 33: ...Blank Page ...
Page 35: ...Blank Page ...
Page 83: ...Blank Page ...
Page 93: ...Blank Page ...
Page 97: ...Blank Page ...
Page 117: ...Blank Page ...