11.9 Selecting the Loading Data Type for the
Communication Command
Procedure
1.
Press the
SHIFT+SETUP (MISC)
keys.
The MISC menu appears.
2.
Press the
Remote I/F
soft key.
The Remote I/F menu appears.
3.
Press the
BinReply
soft key, then select Bin or Asc.
Explanation
Normally this setting is not available. With the channel expansion function, if you have
a GS820 of firmware revision 1.09 or later as a slave device and a GS820 of firmware
revision 1.08 or earlier as the master device, use this setting to specify a data type of
ASCII.
Relationship with the :TRACe:CHANnel<n>:DATA:FORMat command (see page 16-33)
If you use this setting to specify the data type to ASCII, and then use the
:TRACe:
CHANnel<n>:DATA:FORMat
command to set the data format to binary and then load
stored data, the data is loaded as the text string when text is specified.
In all other cases, the stored data is loaded in the data format specified with the
:TRACe:
CHANnel<n>:DATA:FORMat
command.
Note
When using the channel expansion function, it is recommended to use the latest firmware
revision for both the master and slave devices.
<<Corresponding Command Mnemonic>>
:TRACe:BINary:REPLy BINary|ASCii