Explanation
Turning the Limiter ON/OFF
ON: The limiter is activated at the specified limit values.
OFF: The limiter is activated at the boundary of the source range (see section 2.3).
However, the limit values are not displayed.
Tracking
ON: Sets the limit values with the same absolute value but with opposite signs.
Example High limit value: +1.00000 mA
Low limit value: −1.00000 mA
OFF: Set the limit values to arbitrary positive and negative values.
Example High limit value: +1.50000 mA
Low limit value: −1.00000 mA
Setting the Limit Values
In voltage source mode, the current limiter is enabled. In current source mode, the
voltage limiter is enabled.
The optimal limiter range for the specified limit values is automatically selected.
765601/765602
•
Current Limiter
| Setting |
1
Range
Resolution
Minimum Setting
2
10.000 nA to 200.000 nA
200 nA
1 pA
10 nA
0.20001 µA to 2.00000 µA
2 µA
10 pA
10 nA
2.0001 µA to 20.0000 µA
20 µA
100 pA
100 nA
20.001 µA to 200.000 µA
200 µA
1 nA
1 µA
0.20001 mA to 2.00000 mA
2 mA
10 nA
10 µA
2.0001 mA to 20.0000 mA
20 mA
100 nA
100 µA
20.001 mA to 200.000 mA
200 mA
1 µA
1 mA
0.20001 A to 1.20000 A
1 A
10 µA
10 mA
1.20001 A to 3.20000 A
3 A
10 µA
10 mA
1 Larger of the two values |high limit value| or |low limit value| if tracking is OFF
2 Minimum setting if tracking is OFF.
•
Voltage Limiter
| Setting |
1
Range
Resolution
Minimum Setting
2
1.000 mV to 200.000 mV
200 mV
1 µV
1 mV
0.20001 V to 2.00000 V
2 V
10 µV
1 mV
2.0001 V to 7.0000 V
7 V
100 µV
5 mV
7.0001 V to 18.0000 V
18 V
100 µV
5 mV
1 Larger of the two values |high limit value| or |low limit value| if tracking is OFF
2 Minimum setting if tracking is OFF.
5.3 Setting the Limiter
5-6
IM 765601-01E