<4. Setting converter>
4-30
IM 12A01F01-03EN 2nd Edition : Feb. 19, 2019-00
4.7
Error settings
When an error occurs, the screen of the converter flashes on until the error is cancelled.
You can select and configure diagnostic notification of error status according to NE107: “Failure”,
“Out of specification”, “Function check”, “Maintenance required” or “Off”.
When “Alarm NE107”* is “On”, (“On” as factory default), the icons defined in NAMUR NE107*,
one of
, , ,
lights on upon detection of each alarm, on Main screen or Home screen.
If you set “Alarm NE107” to “Off”, only
(Fault) or
(Warning) flashes on Main/Home screen.
Status
Initial
Description
Alarm
NE107
Burn
out
On
Off
Failure
F
Signal invalid due to malfunction in the device, sensor or actuator
Yes*
Function
check
C
Signal temporarily invalid due to on-going work on the device.
No
Out of
Specification
S
Permissible ambient or process conditions exceeded or the
measuring uncertainty of sensors or deviations from the set value
in actuators is probably greater than expected.
Maintenance
Required
M
Although the signal is valid, the remaining life is nearly exhausted
or a function will soon be restricted due to operational conditions.
*: “Failure” activates Burn. If you set “Burn” to “Off” at 4.3.1
Output, fail is only displayed on screen.