<3. Converter menu>
3-7
IM 12A01F01-03EN 2nd Edition : Feb. 19, 2019-00
3.4
Reset
If you need to reset to factory default, follow this procedure. In a usual operation, you don’t need
reset.
CAUTION
Once you reset the converter, every data of, language, password, device setting is cleared and
reset to default, excluding date/time.
If you want to reset after configuring some settings, store the data into the memory of FLXA402, if
necessary. See 5.1.3 Save configuration
To connect SENCOM SA, display the measurement or establish communication of HART or mA
output, you need to set up network system for SENCOM SA and assign each sensor-module
address of HMI.
l
How to reset parameters
1.
On Home/Main screen, tap to go to Converter menu.
2. Press
“Reset”.
3.
On “Load factory defaults”, find the sensor module type in use and select.
Select the appropriate connection to suit your sensors from the list below.
List of Load factory defaults
PH
SC
ISC
DO
PH+PH
SC+PH
ISC+PH
DO+PH
PH+SC
SC+SC
ISC+SC
DO+SC
PH+ISC
SC+ISC
ISC+ISC
DO+ISC
PH+DO
SC+DO
ISC+DO
DO+DO
4.
Confirm the load selection you choose on the step 3. (above) is displayed and tap
“Execute”.
(1)
(2)
(3)
(4)
Figure 3.7
How to reset parameters