HART Communication
7
TI 12A01F01-61EN
Nov. 20, 2018-00
HART Status Group CMD48
byte
bit
HART Alarm Name
Alarm
Number NE107
Device Specific Status 4
4
0 Simulated mA value 1
033
C
1 Simulated mA value 2
034
C
2 Simulated mA value 3
035
C
3 Simulated mA value 4
036
C
4 Outputs in HOLD1
037
N
5 Outputs in HOLD2
038
N
6 Outputs in HOLD3
039
N
7 Outputs in HOLD4
040
N
Device Specific Status 5
5
0
Output configuration error
041
F
1
Other configuration error
042
C
2 Reserved
043
N
3 Reserved
044
N
4 Reserved
045
N
5 Reserved
046
N
6 Reserved
047
N
7 Reserved
048
N
Extended Device Status
6
0 Maintenance Required(NE107: M)
-
N
1 Device Variable Alert *1
-
S
2 Critical Poew Failure
F
3 Failure (NE107: F)
-
N
4
Out of Specification (NE107: S)
-
N
5 Function Check (NE107: C)
-
N
6 Reserved
N
7 Reserved
N
Device Operationg Mode
7
0 Reserved
N
1 Reserved
N
2 Reserved
N
3 Reserved
N
4 Reserved
N
5 Reserved
N
6 Reserved
N
7 Reserved
N
Standardized Status0
8
0 Device Variable Simulation Active
C
1 Non-Volatile Memory Defect
F
2 Volatile Memory Defect
F
3
Watchdog Reset Executed
F
4 Power Supply Conditions Out of Range
S
5 Environmental Conditions Out of Range
S
6 Electronic Defect
F
7
Device Configuration Locked
-
N
Standardized Status1
9
0 Status Simulation Active
-
N
1 Discrete Variable Simulation Active
-
C
2
Event Notification Overflow
N
3 Reserved
N
4 Reserved
N
5 Reserved
N
6 Reserved
N
7 Reserved
N
Analog Channel
Saturated
10
0 Analog Channel 1
N
1 Analog Channel 2
N
2 Analog Channel 3
N
3 Analog Channel 4
N
4 Reserved
N
5 Reserved
N
6 Reserved
N
7 Reserved
N