16-2
IM DLM3054-01EN
Measurement (Measure Setup)
Mode (Mode)
OFF: Disables measurement.
Param: Measures the value of the selected parameter and performs cursor measurement.
Cursor Measurement (Cursor1 and Cursor2)
Measurement items C1, C2, and ΔC are used to measure the Cursor1 value, the Cursor2 value, and the
difference between the two cursor values. The selectable range varies depending on the source axis as follows:
• If the source axis is vertical (Cursor1 =, Cursor2 =): In the range of ±4 divisions of the vertical axis
• If the source axis is horizontal (Cursor1 II, Cursor2 II): In the range of ±4 divisions of the time axis
Measurement Items (Item)
The instrument measures the values of the items listed below that you select.
Peak
Peak value
Max
Maximum value
Min
Minimum value
Mean
Mean value
σ
Standard deviation of the histogram
Median
Median*
Integ±σ
Percentage of values that fall within ±σ (%)
Integ±2σ
Percentage of values that fall within ±2σ (%)
Integ±3σ
Percentage of values that fall within ±3σ (%)
C1
Cursor1 value
C2
Cursor2 value
ΔC
Difference between Cursor1 and Cursor2
* Re-sorts the sampled points from the minimum to the maximum value and determines the value of the middle
number.
σ
±
σ
±3
σ
Peak
Mean
±2
σ
Max
Min
16 Waveform Histogram Display