15-30
IM DLM3054-01EN
List Display (List)
The list displays the following items.
When Grouping Is Set to ON
No.
Analysis number. Negative numbers are assigned to frames before the and positive numbers
are assigned to frames after the trigger position.*
The instrument can display the analysis results for up to 300000 frames in the range of
−299999 to 299999. Pressing the RESET key highlights data number zero.
Time(ms)
Displays the time from the trigger position* to the start of the byte in milliseconds.
Data
Displays the first 16 bytes of data for each group in hexadecimal notation and ASCII codes. You
can press the Detail soft key to view all of the data of the specified group (analysis number).
Information
Displays the following error information. If multiple errors are detected in one piece of data, the
instrument only displays the framing error indication.
Framing Error, Parity Error
* If
is set to OFF, they are based on the trigger point.
When Grouping Is Set to OFF
Addr
Displays the address relative to the first data address.
Hex
Displays the data in hexadecimal notation. An asterisk or an “X” is attached to the data if a
framing error or a parity error is detected, respectively.
Ascii
Displays the data using ASCII codes.
Search Setup (Search)
Search Type (Mode)
Select the UART signal search type from one of the settings below.
• Every Data: Searches for data stop bits
• Error: Searches for errors
• Data: Searches for data patterns
The method to set search conditions is the same as that for the trigger conditions of the UART trigger.
Executing a Search (Search)
The instrument searches for locations where the specified search conditions are met. Then, the instrument
displays the specified detected point number (Pattern No) expanded in the zoom window.
Detected Points
The detected points are the same as the trigger points.
15 Analyzing and Searching Serial Bus Signals