1-5
IM DLM3054-01EN
Unit (Unit)
You can set the unit using up to four alphanumeric characters.
Scaling Coefficient (A) and Offset (B)
The selectable ranges of scaling coefficient A and offset B are as follows:
Selectable of A and B
−10.000E+30 to +10.000E+30
Default settings
A
1.0000E+00
B
0.0000E+00
Label Display (Label)
You can assign labels using up to 8 characters to be displayed.
Turning Label Display On and Off (Display)
• ON: Shows labels
• OFF: Hides labels
Label (Name)
Set the label name when you set the label display to ON.
Bandwidth Limit (Bandwidth)
You can set a upper bandwidth limit on the analog signal for each channel.
This allows you to observe waveforms with components above the specified frequency removed.
Select the frequencies to limit from the following.
FULL (no bandwidth limit), 200 MHz, 100 MHz, 20 MHz, 10 MHz, 5 MHz, 2 MHz, 1 MHz, 500 kHz, 250 kHz, 125
kHz, 62.5 kHz, 32 kHz, 16 kHz, 8 kHz
When high resolution mode is on, the bandwidth limit is set at 200 MHz even when bandwidth limit is set to
FULL.
Offset (Offset)
If you set an offset when measuring an analog signal that is riding on top of a given voltage (or current), the
given voltage (or current) is subtracted from the input signal. This allows you to only view the changes in the
signal at a higher vertical scale setting.
You can set the offset for each channel.
Selectable Range for Voltage
Input Coupling
Voltage Scale (With probe = 1:1)
Selectable Offset Voltage Range
AC/DC
500 μV/div to 50 mV/div
−1.0 V to 1.0 V
0.1 V/div to 0.5 V/div
−10.0 V to 10.0 V
1 V/div to 10 V/div
−100.0 V to 100.0 V
DC50
500 μV/div to 50 mV/div
−1.0 V to 1.0 V
0.1 V/div to 1 V/div
−5.0 V to 5.0 V
You can set the offset in 0.01 mV steps. For 500 μV/div, you can set the offset in 5 μV steps.
Resetting the Offset
Press the RESET key to set the offset value to 0 V.
1 Vertical Axis (Analog Signal)