SM 700820-01E
2 - 10
3) Connect the function generator output to the CH1 input.
Acquire data with the slope in RISE and the trigger level at 300 mV, read the trigger
point data ten times using the cursor, and check that the average value is within the
following range: 170 mV
≤
Cursor1
≤
430 mV
Similarly, each average value of ten trigger point data should be within the following
corresponding range:
Slope
Trigger level
Permissible range
300 mV
170 mV
≤
Cursor1
≤
430 mV
300 mV
170 mV
≤
Cursor1
≤
430 mV
0 mV
–100 mV
≤
Cursor1
≤
100 mV
0 mV
–100 mV
≤
Cursor1
≤
100 mV
–300 mV
–430 mV
≤
Cursor1
≤
–170 mV
–300 mV
–430 mV
≤
Cursor1
≤
–170 mV
4) Test all channels in the same manner.
2.10 Logic Input Function Test
Equipment required
Equipment
Qty
Mandatory specifications
Recommended
Probe
1
150 MHz frequency range
Yokogawa 700998
Logic Probe
4
8 bit
Yokogawa 700985
Connection
Logic Probes
Input (Red)
CAL
Gnd. (Black)
DL2700
Procedure
1) Turn on the power of the DL2700 and press the INITIALIZE key, then press the
“Initialize” soft key to initialize the settings.
2) Press the AUTO SETUP key and select the “Auto Setup” option.
3) Press the MISC key , select the “Logic” option.
4) Press the SELECT key and choose the “Logic A” option
5) Press the ACQ SELECT and select the “POD” option
6) Press the Threshold Level and select the “User” option
7) Turn the jog shuttle to adjust the “User ThrLev” to 0.5V.
8) Set the POD B to POD D by the same way as POD A.
The ACQ SELECT menu is appeared according to analog input number.