5-75
IM 701730-17E
Commands
5
:MEASure:CONSt<x>
Function
Sets the constants to be used in the automated
measurement of waveform parameters of Dual
Area or queries the current setting.
Syntax
:MEASure:CONSt<x> {<NRf>}
:MEASure:CONSt<x>?
<NRf>=–9.9999E+30 to 9.9999E+30
<x>=1 to 4
Example
:MEASURE:CONST1 1.0000E+00
:MEASURE:CONST1? ->
:MEASURE:CONST1 1.0000E+00
:MEASure:CYCLe?
Function
Queries all settings related to the periodic
statistical processing of the automated
measurement of waveform parameters (Cycle
Statistics).
Syntax
:MEASure:CYCLe?
Example
:MEASURE:CYCLE? ->
:MEASURE:CYCLE:TRACE 1
:MEASure:CYCLe:ABORt
Function
Aborts the periodic statistical processing of the
automated measurement of waveform
parameters (Cycle Statistics).
Syntax
:MEASure:CYCLe:ABORt
Example
:MEASURE:CYCLE:ABORT
:MEASure:CYCLe:EXECute
Function
Executes the periodic statistical processing of
the automated measurement of waveform
parameters (Cycle Statistics).
Syntax
:MEASure:CYCLe:EXECute
Example
:MEASURE:CYCLE:EXECUTE
:MEASure:CYCLe:TRACe
Function
Sets the waveform used to determine the
period for the periodic statistical processing of
automated measurement of waveform
parameters (Cycle Statistics) or queries the
current setting.
Syntax
:MEASure:CYCLe:TRACe {OWN|<NRf>|
MATH<x>}
:MEASure:CYCLe:TRACe?
<NRf>=1 to 4 (1 to 2 on the DL1720E)
<x>=1 or 2 (1 on the DL1720E)
Example
:MEASURE:CYCLE:TRACE 1
:MEASURE:CYCLE:TRACE? ->
:MEASURE:CYCLE:TRACE 1
:MEASure:HISTory:ABORt
Function
Aborts the statistical processing within the
historical data of the automated measurement
of waveform parameters (History Statistics).
Syntax
:MEASure:HISTory:ABORt
Example
:MEASURE:HISTORY:ABORT
:MEASure:HISTory:EXECute
Function
Executes the statistical processing within the
historical data of the automated measurement
of waveform parameters (History Statistics).
Syntax
:MEASure:HISTory:EXECute
Example
:MEASURE:HISTORY:EXECUTE
:MEASure:MODE
Function
Sets the mode of the automated measurement
of waveform parameters or queries the current
setting.
Syntax
:MEASure:MODE {OFF|ON|STATistics|
CYCLE|HISTory}
:MEASure:MODE?
Example
:MEASURE:MODE ON
:MEASURE:MODE? -> :MEASURE:MODE ON
:MEASure:ONECycle
Function
Turns ON/OFF the automated measurement in
the one-cycle mode or queries the current
setting.
Syntax
:MEASure:ONECycle {<Boolean>}
:MEASure:ONECycle?
Example
:MEASURE:ONECYCLE ON
:MEASURE:ONECYCLE ->
:MEASURE:ONECYCLE 1
:MEASure:TLINkage<x>
Function
Turns ON/OFF the linked movement of the
cursors in the automated measurement of
waveform parameters or queries the current
setting.
Syntax
:MEASure:TLINkage<x> {<Boolean>}
:MEASure:TLINkage<x>?
<x>=1, 2
Example
:MEASURE:TLINKAGE1 ON
:MEASURE:TLINKAGE1 ->
:MEASURE:TLINKAGE1 1
Description
“
:MEASURE:TLINKAGE1
” sets the links for T-
Range1 and T-Range2 or queries the current
settings. “
:MEASURE:TLINKAGE2
” sets the
links for T-Range3 and T-Range4 or queries the
current settings.
:MEASure:TRANge<x> (Time Range)
Function
Sets the measurement range of the automated
measurement of waveform parameters or
queries the current setting.
Syntax
:MEASure:TRANge<x> {<NRf>,<NRf>}
:MEASure:TRANge<x>?
<x>=1, 2
<NRf>=–5 to 5 div (10 div/displayed record
length steps)
Example
:MEASURE:TRANGE1 -4,4
:MEASURE:TRANGE1? ->
:MEASURE:TRANGE1 -4.00E+00,4.00E+00
Description
“
:Measure:TRANge1
” sets the measurement
range of Area1 during normal measurement
and during measurement on dual areas or
queries the current setting.
“
:Measure:TRANge2
” sets the measurement
range of Area2 during the measurement on
dual areas or queries the current setting.
5.17 MEASure Group