7-7
IM 701510-01E
Changing
Acquisition and Display Conditions
7
7.4
Acquiring Waveforms using the Sequential
Store Function (for DL1520L/DL1540/DL1540L)
Acquiring the waveform data only a specified number of times
1.
Press the
MODE key
to display the trigger mode setting menu.
2.
Press the “
N-SGL
” soft key. When a trigger is activated at the time the
“
N-SGL
” soft key is pressed, waveform acquisition starts.
For DL1520L/DL1540L, “
SINGLE
” menu will be displayed instead of
“
SGL(S)
” and “
SGL(L)
” menus.
3.
Turn the rotary knob to set the number of acquisitions at
“
Acquisition Count (N)
”. When waveform acquisition is already in
progress and you change the number of acquisitions, make sure to
restart to get the specified number of acquisitions.
4.
If no trigger has been activated already, press the
START/STOP key
or once more the “
N-SGL
” soft key to start data acquisition.
Displaying a specific acquired data by selecting its number
5.
Press the
CLEAR TRACE key
to clear the displayed waveforms.
6.
Press the
DISPLAY key
to display the display selection menu.
7.
Use the rotary knob to set the number of the data you want to see. The
corresponding waveform will appear at the same time.
Displaying a specific acquired data by selecting its time of
acquisition
8.
Continuing from step 6, press the “
Time Stamp
” soft key to display
the list of acquired data with their acquisition times.
9.
Use the rotary knob to highlight the data you want to see.
10.
Press the
SELECT key
. The list will disappear and the selected
waveform will appear on the screen.
Keys and Procedure
MODE key
Rotary knob
Soft keys
START/STOP key
CLEAR TRACE key