<APPENDIX 3. OPERATION OF EACH PARAMETER IN FAILURE MODE>
A3-9
IM 01F06F00-01EN
LCD
Display Alarm Detail
DI1 Block
DI2 Block
PID Block
IT Block
AR Block
Alarm Reset SW*
(default)
AL-27 PID in O/S
Mode (AL-27)
<BLOCK_ERR>
Out of Service
Provided
(OFF)
<OUT.Status>
Bad-Out of Service
AL-28 AI3 in O/S
Mode (AL-28)
Provided
(OFF)
AL-29 IT in O/S Mode
(AL-29)
<BLOCK_ERR>
Out of Service
Provided
(OFF)
<OUT.Status>
Bad -Out of Service
AL-30 AR in O/S
Mode (AL-30)
<BLOCK_ERR>
Out of Service
Provided
(OFF)
<OUT.Status>
Bad-Out of Service
AL-41 Flow Rate Over
Range (AL-41)
<PV_D.Status>
• TARGET in TB’s LIMSW = PRIMARY_
VALUE
Uncertain-Non Specific
Not provided
<OUT_D.Status>
• TARGET in TB’s LIMSW = PRIMARY_
VALUE
Uncertain-Non Specific
AL-42
Flow Span
Exceed Limit
(AL-42)
<PV_D.Status>
• TARGET in TB’s LIMSW = PRIMARY_
VALUE
Uncertain-Non Specific
Not provided
<OUT_D.Status>
• TARGET in TB’s LIMSW = PRIMARY_
VALUE
Uncertain-Non Specific
AL-43 Temp. Over
Range (AL-43)
<PV_D.Status>
*1
• TARGET in TB’s LIMSW =
SECONDARY_VALUE
Uncertain-Non Specific
*2
Uncertain-Non Specific
Not provided
<OUT_D.Status>
*1
• TARGET in TB’s LIMSW =
SECONDARY_VALUE
Uncertain-Non Specific
*2
Uncertain-Non Specific
AL-51
Transient
Vibration (AL-
51)
<PV_D.Status>
• TARGET in TB’s LIMSW = PRIMARY_
VALUE
Uncertain-Non Specific
Provided
(OFF)
<OUT_D.Status> (OFF)
• TARGET in TB’s LIMSW = PRIMARY_
VALUE
Uncertain-Non Specific
AL-52 High Vibration
(AL-52)
<PV_D.Status>
• TARGET in TB’s LIMSW = PRIMARY_
VALUE
Uncertain-Non Specific
Provided
(OFF)
<OUT_D.Status> (OFF)
• TARGET in TB’s LIMSW = PRIMARY_
VALUE
Uncertain-Non Specific