61
IM 04L41B01-02E
Setup Items and
Default V
a
lues
Setup Item
Selectable Range or Selections
Default Value
Setting
Mode=1-5V
Range
1-5V
1-5V
Span Lower
0.800 to 5.200
1.000
Span Upper
0.800 to 5.200
5.000
Scale Lower
–30000 to 30000, decimal position :0 to 4
0.00
Scale Upper
–30000 to 30000, decimal position :0 to 4
200.00
Unit
6 characters or less
–
Low-cut
On/Off
Off
Mode=Sqrt
Range
20mV/60mV/200mV/2V/6V/20V/50V
Span Lower
Depends on the range.
–
Span Upper
Depends on the range.
–
Scale Lower
–30000 to 30000, decimal position: 0 to 4
0.00
Scale Upper
–30000 to 30000, decimal position: 0 to 4
200.00
Unit
6 characters or less
–
Low-cut
On/Off
Off
Low-cut value
0.0 to 5.0
0.5
Alarm
1, 2, 3, 4
On/Off
Off
Type
H:High/L:Low/h:delta H/l:delta L/R:rate H/
H
r:rate L/T:delayH/t:delayL
Value
Numerical value
–
Relay
On/Off
Off
Number
I01/I02/.../I06 (Depends on the model.)
I01
S01/S02/S03/.../S29/S30
Detect
On/Off
On
Meas channel > Tag, Memory, Delay
Setup Item
Selectable Range or Selections
Default Value
Setting
First-CH, Last-CH
1/2/3/.../11/12 (Depends on the model.)
1
Tag > Characters
16 characters or less
–
Memory sample > On/Off
On/Off
On
Alarm delay > Time
1 to 3600s
10
Meas channel > Moving average
Setup Item
Selectable Range or Selections
Default Value
Setting
First-CH, Last-CH
1/2/3/.../11/12 (Depends on the model.)
1
Moving average > On/Off
On/Off
Off
Moving average > Count
2 to 400
2
Meas channel > Color
Setup Item
Selectable Range or Selections
Default Value
Setting
Group of channel
001-006, etc
Depends on the model.
Color
Red/Green/Blue/B.violet/Brown/Orange/
Red to Gray
Y.green/Lightblue/Violet/Graly/Lime/Cyan/
Darkblue/Yellow/Lightgray/Purple/Black/Pink/
L.brown/L.green/Darkgray/Olive/DarkCyan/
S.green (24 colors)
Meas channel > Zone, Scale
Setup Item
Selectable Range or Selections
Default Value
Setting
First-CH, Last-CH
1/2/3/.../11/12 (Depends on the model.)
1
Zone > Lower
0 to 95
0
Zone > Upper
5 to 100
100
Scale > Position
Off/1/2/3/4/5/6
1
Scale > Division
4/5/6/7/8/9/10/11/12/C10
10
Setup Items and Default Values