SM 04L01A01-01E
4-4
4.2 AD Board Offset and Gain Adjustment
8
If all ranges are set, push the ESC key. The screen returns to figure 4.3.
Select item #3 to end the task.
9
The dialog box in figure 4.5 appears.
Select Yes to save the calibrated value to the EEPROM. The screen will return to
figure 4.2.
Figure 4.5 calibration value saving screen.
10 Repeat steps 3 to 9 for all A/D number’s.
11 If adjustment was successful turn the power to the UUA off.
When you select task #1,
Display
in figure 4.3, the screen below (figure 4.6) will appears.
Confirm the calibration value of each range (decimal value = 215: shows converted 15
bits data.)
Figure 4.6 Calibration value confirmation screen.
1
When confirmation is finished, press ESC to return to the screen in figure 4.3.
Select item #3 to end the task.
2
After step 1, the dialog box in figure 4.5 will appear.
Select No for normal operation.
CAUTION
Do not change the displayed value, as it influences the measured value.