22
10. Specifications
10.1 General Specifications
Display count:
9999 / 6000
Measuring rate:
3 times / sec.
Over range indicator:
“OL” or “-OL”
Auto Power Off:
Approx. 15 minute.
Low-battery indicator:
(four steps)
Power supply:
9V alkaline battery (6LR61)
Battery life:
When using alkaline battery, backlight off
Approx. 20 hours
Operating temperature and humidity:
0 to 50 °C (with no condensation)
≤ 80% RH (0 to 30 °C)
≤ 75% RH (30 to 40 °C)
≤ 45% RH (40 to 50 °C)
Temperature coefficient: At 0 to 18 °C and 28 to 50 °C
Add 23±5 °C accuracy × 0.2 / °C
Storage temperature:
-10 to 50 °C, 80% RH or less (remove the battery)
Withstand voltage:
AC 6880 Vrms 5 seconds (between the core and the case)
AC 4300 Vrms 5 seconds
(between the core and the voltage input terminals)
AC 6880 Vrms 5 seconds
(between the voltage input terminals and the case)
Insulation resistance:
100 MΩ or greater at 1000 VDC
(between the core and the case,
the core and the voltage input terminals and
the voltage input terminals and the case)
Safety standards:
EN61010-1, EN61010-2-033, EN61010-2-032
1000 V CAT III, 600 V CAT IV
EN61010-031 (the test leads)
Pollution degree 2, indoor use,
altitude 2000m or less
EMC standards:
EN61326-1 Class B,
EN61326-2-1, EN61326-2-2,
EMC Regulatory Arrangement in Australia and
New Zealand
EN55011 Class B Group 1
Korea Electromagnetic Conformity Standard
(
한국 전자파적합성기준)
Environmental standard
:
EN50581 Monitoring and control instruments
Dimensions:
Approx. 87.5 mm (W) × 242 mm (L) × 51 mm (D)
Diameter of measurable conductors: 37mm (Maximum)
Weight:
Approx. 435g (including the battery)
Accessories:
Test leads 1set (Red and Black)
Soft case: 93044
9V alkaline battery (6LR61)
User’s Manual