YOKOGAWA CL360 User Manual Download Page 6

IM CL360

 13

 

    Analogue Output (Output impedance: Approx. 1k

)

 

  AC Output 

Range 

Measuring range 

AC Output Voltage 

Accuracy 

200mA 0 

to 

200mA 

2A 0 

to 

2A 

20A 0 

to 

20A 

±2.0% rdg 

200A 0 

to 

200A 

0 to 200mV 

±2.5% rdg 

0 to 500A 

0 to 50mV 

±3.0% rdg 

1000A 

501 to 1000A 

50 to 100mV 

±5.0% rdg 

Voltage proportional to the current under test is output with “WIDE” frequency 
characteristics regardless of the setting of the frequency selector or peak hold switch. 

 
  DC Output 

Accuracy (Frequency Range) 

Range  Measuring range 

DC Output 

Voltage 

Freq. Selector 

Switch at WIDE 

position 

Freq. Selector 

Switch at 50/60Hz 

position 

200mA 0 

to 

200mA 

2A 0 

to 

2A 

20A 0 

to 

20A 

±3.0% rdg 

±3.5% rdg 

200A 0 

to 

200A 

0 to 200mV 

±3.5% rdg 

±4.0% rdg 

0 to 500A 

0 to 50mV 

±5.0% rdg 

±5.5% rdg 

1000A 

501 to 1000A 

50 to 100mV  ±7.0% rdg 

±7.5% rdg 

 

„

 General Specifications 

  Operating System : Dual integration 

  Measurement Function : AC current 

  Display : Field effect 3 - 1/2 digit liquid crystal display with maximum count 

of 1999 

  Range selection : Manual 

  Overrange Indication : "1" is displayed on the highest digit except for 

1000A AC range. 

  Response Time : Approx. 1 seconds. 

  Sample Rate : Approx. 3 times per second. 

  Temperature and Humidity for Guaranteed Accuracy : 23°C ±10°C, relative 

humidity up to 85% without condensation 

  Operating Temperature and Humidity : 0 to 40°C, relative humidity up to 

85% without condensation 

14 

IM CL360

  Storage Temperature and Humidity : -10 to 50°C, relative humidity up to 

80% without condensation 

  Effect of conductor position: Within 2% difference between maximum and 

minimum values to a 10 mm-dia conductor, at every part inside the jaws 

  Effect of External Magnetic Field :   

15mA AC max. in proximity to a 10mm-dia conductor carrying 100A AC. 

 Effect of Residual Current : 10mA AC max. when clamping on two 

10mm-dia conductors, each carrying supply or return 100A AC current. 

  Power Source : 6LR61 or 6F22 DC9V battery 

  Battery Life : Approx. 60 hours (continuity) 

  Current Consumption : Approx. 5mA max. 

  Withstanding Voltage : 3700V AC, 50/60Hz for 1 minute between electrical 

circuit and housing case or metal part of the jaws 

  Insulation Resistance : 10M

 or greater at 1000V between electrical circuit 

and housing case or metal part of the jaws 

  Conductor Size : Approx. 68mm diameter max. 

  Dimensions : Approx.130(W) x 250(H) x 50(D) mm 

  Weight : Approx. 570g 

  Safety Standard: EN 61010-1 

 EN 

61010-2-032 

AC 300V CAT III, AC 600V CAT II,   
Pollution degree 2, indoor use 

  EMC Standard:  EN 61326 

 EN 

55022 

 Accessories : 6F22 battery······························ 1 

Carrying case Model 93031 ······ 1 
User’s Manual ··························· 1 

  Optional Accessories : Output cable with banana plug    Model 91020 

  Peak Hold : Response time selectable from approx. 10ms and 100ms. 

1/

2

  of the peak current is displayed, which means an RMS reading will 

be obtained when current having sinusoidal waveform in measured. 
Peak hold reading varies by approximately ±1% max of full scale in the 
first minute at 23°C ±10°C and relative humidity of 75% without condensa-
tion. 

 

IM CL360

 15

 

5.  Calibration and After-sales Service 

Should any failure occur while you are using the tester, follow the instructions 
given below. If the tester still fails to operate correctly and needs repair, con-
tact the vendor from whom you purchased the instrument or the nearest 
Yokogawa M&C sales office. 
 

Condition Possible 

Cause 

Remedy 

Display blanks after 
power-on 

Battery is improperly installed.
Battery is exhausted. 

Install the battery correctly. 
Replace the buttery. 

Display reading 
remains frozen. 

Data hold button is pressed in.
Peak hold switch is at ON   
position. 

Release data hold button. 
Set peak hold switch to OFF 
position. 

Transformer jaws buzz 
when measuring large 
current. 

----------------- 

This is not a failure 

The lowest digit of 
reading is unstable 

----------------- 

This is not a failure. The 
instrument is highly accu-
rate, so it senses slight 
variations in the current 
under test. 

Output cable 
(Model 91020) does not 
output voltage. 

The cable is open circuit. 

Check the cable for an open 
circuit. 

 
Calibration  
It is recommended that the instrument be calibrated once every year. 

Summary of Contents for CL360

Page 1: ...instrument if the operating procedure is not followed correctly and describes the precautions for avoiding such injury or damage NOTE Draws attention to information essential for understanding the op...

Page 2: ...res or dew fall Never apply voltage to the OUTPUT terminal The terminal is not electri cally isolated from the internal circuits of the instrument Use a damp cloth and detergent for cleaning the instr...

Page 3: ...t 1 Set the range switch to the desired position Do not exceed the maximum allowable input current for the selected range 2 For normal measurement press the open close lever to open the trans former j...

Page 4: ...Time 100ms Peak Response Time Input Waveform Hold Capacitor Voltage 100 90 100 90 10ms t 100ms t For instance select the 10ms response time when measuring a surge current that will occur when a power...

Page 5: ...utput DC output Recorder Oscilloscope connect to oscilloscope or multimeter Model CL360 Multimeter OFF IM CL360 11 3 Battery Replacement WARNING To avoid electric shock hazard make sure to set the Ran...

Page 6: ...hours continuity Current Consumption Approx 5mA max Withstanding Voltage 3700V AC 50 60Hz for 1 minute between electrical circuit and housing case or metal part of the jaws Insulation Resistance 10M o...

Page 7: ...ERING INSTRUMENTS KOREA CORPORATION KOREA Phone 82 2 551 0660 to 0664 Facsimile 82 2 551 0665 YOKOGAWA AUSTRALIA PTY LTD AUSTRALIA Phone 61 2 9805 0699 Facsimile 61 2 9888 1844 YOKOGAWA BLUE STAR LTD...

Reviews: