YOKOGAWA CL235 User Manual Download Page 6

 

IM CL235

 13

 

5. Specifications 

„

 Instrument Specifications 

 Measuring Ranges and Accuracy (at 23±5°C, 45 to 75% relative humidity) 

  DC Current 

Measuring Range (Auto-ranging) 

Accuracy 

0 to 399.9A 
150 to 1000A 

±1.0% rdg ±5dgt 

      Auto-ranging 
        Lo : 0 to 399.9A (Shifts to Hi at 399.9A) 
        Hi : 150 to 1000A (Shifts to Lo at 150A. At 1020A or above, "OL" is shown.) 

 

  AC Current 

A (Crest factor (CF): 3.0) 

Measuring Range (Auto-ranging) 

Accuracy 

0 to 399.9A 
150 to 600A 

±1.5% rdg ±5dgt (50/60Hz) 
±3.5% rdg ±5dgt (40 to 1kHz) 

      Conversion method : AC coupled, true rms responding, calibrated to the rms 
      Auto-ranging 
        Lo : 0 to 399.9A (Shifts to Hi at 399.9A) 
        Hi : 150 to 600A (Shifts to Lo at 150A. At 620A or above, "OL" is shown.) 

 

  DC Voltage 

V (Input impedance: 2M

Measuring Range (Auto-ranging) 

Accuracy 

0 to 39.99V 
15.0 to 399.9V 
150 to 600V 

±1.0% rdg ±5dgt 

      Auto-ranging 
        Lo : 0 to 39.99V (Shifts to Mid at 39.99V) 
        Mid : 15.0 to 399.9V (Shifts to Lo at 15.0V and to Hi at 399.9V) 
        Hi : 150 to 600V (Shifts to Mid at 150V. At 620V or above, "OL" is shown.) 

 

  AC Voltage 

V Input impedance: 2M

 (Crest factor (CF): 3.0 or less) 

Measuring Range (Auto-ranging) 

Accuracy 

0 to 39.99V 
15.0 to 399.9V 
150 to 600V 

±1.5% rdg ±5dgt (50/60Hz) 
±3.5% rdg ±5dgt (40 to 1kHz) 

      Conversion method : AC coupled, true rms responding, calibrated to the rms 
      Auto-ranging 
        Lo : 0 to 39.99V (Shifts to Mid at 39.99V) 
        Mid : 15.0 to 399.9V (Shifts to Lo at 15.0V and to Hi at 399.9V) 
        Hi : 150 to 600V (Shifts to Mid at 150V. At 620V or above, "OL" is shown.) 

 

14 

IM CL235

  Resistance 

 

Measuring Range (Auto-ranging) 

Accuracy 

0 to 399.9

 

150 to 3999

 

±1.0% rdg ±5dgt 

      Auto-ranging 
    Lo : 0 to 399.9

 (Shifts to Hi at 400.0

    Hi : 150 to 3999

 (Shifts to Lo at 150

. At 3999

 or above, "OL" is shown.) 

 

  Continuity Check 

 

Measuring Range 

Accuracy 

0 to 399.9

 

±1.0% rdg ±5dgt 

      The buzzer beeps when the resistance is 20

 or less. 

 

  Frequency 

Measuring Range 

Accuracy 

10 to 3000Hz 

±1.5% rdg ±5dgt 

 

 

IM CL235

 15

 

*Effective Value (rms) 

Most alternating currents and voltages are expressed in effective values, 
which are also referred to as RMS (Root-Mean-Square) values. 
The effective value is the square root of the average of square of alternating 
current or voltage values. 
Many clamp meters using a conventional rectifying circuit have "RMS" scales 
for AC measurement. The scales are, however, actually calibrated in terms of 
the effective value of a sine wave though the clamp meter is responding to 
the average value. The calibration is done with a conversion factor of 1.111 
for sine wave, which is found by dividing the effective value by the average 
value. These instruments are therefore in error if the input voltage or current 
has some other shape than sine wave. 

Waveform

Effective

value

Average

value

Conversion

factor

Reading

errors for

average

sensing

instruments

Crest factor

 

      *CF(Crest Factor) is found by dividing the peak value by the effective value. 
   Examples: 

DC: CF=1 
Sine wave: CF=1.414 
Square wave with a 1:10 duty ratio: CF=3 

 

16 

IM CL235

„

 General Specifications 

  Operating System : Dual integration 

  Measurement Function : AC current, DC current, AC voltage, DC voltage, 

resistance, continuity check, frequency 

  Display : Liquid crystal display with maximum counts of 3999 

  Overrange Indication : "OL" is shown on the display 

  Response Time : Approx. 2 seconds. 

  Sample Rate : Approx. 3 times per second. 

  Temperature and Humidity for Guaranteed Accuracy : 23°C ±5°C, relative 

humidity up to 75% without condensation 

  Operating Temperature and Humidity : 0 to 40°C, relative humidity up to 

90% without condensation 

  Storage Temperature and Humidity : -10 to 50°C, relative humidity up to 

75% without condensation 

  Effect of conductor position : Within 2% difference between maximum and 

minimum values to a 10 mm-dia conductor, at every part inside the jaws 

  Power Source : 6F22 or 6LR61 9V battery 

  Battery Life : Approx. 15 hours (continuity) 

  Current Consumption : Approx. 15mA max. 

  Sleep function : Automatically powered down in Approx. 30 minutes after 

the last switch operation (power consumption : Approx. 20

µ

A) 

  Withstanding Voltage : 5500V AC, 50/60Hz for 1 minute between electrical 

circuit and housing case or metal part of the jaws 

  Insulation Resistance : 10M

 or greater at 1000V between electrical circuit 

and housing case or metal part of the jaws 

  Conductor Size : Approx. 33mm diameter max. 

  Dimensions : Approx. 91(W) x 208(H) x 40(D) mm 

  Weight : Approx. 450g (with battery) 

  Safety Standard : IEC61010-1:1995 

IEC61010-2-032:1994 
AC/DC 600V CAT III, Pollution degree 2, indoor use 

  EMC Standard : BS EN 55022:1995 

  Radiation immunity : IEC801-3:1993 

  Accessories : Test leads Model 98010 ············ 1set 

6F22 or 6LR61 battery ·············· 1 
Carrying case Model 93032 ······ 1 
User’s Manual ··························· 1 

  Optional Accessories : Clamp adapter Model 99025 

 

Summary of Contents for CL235

Page 1: ... Indicates that there is a possibility of serious personal injury of damage to the instrument if the operating procedure is not followed correctly and describes the precautions for avoiding such injury or damage NOTE Draws attention to information essential for understanding the op eration and features ii IM CL235 WARNING Never make measurement on a circuit above 600V AC or 600V DC Do not use the ...

Page 2: ...h to the OFF position after use When the instrument will not be in use for a long period of time Place it in storage after removing the battery Use a damp cloth and detergent for cleaning the instrument Do not use abrasives or solvents NOTE Radiation immunity affects the accuracy of CL235 testers under the condi tions specified in IEC801 3 1993 If equipment generating strong electromagnetic interf...

Page 3: ...t WARNING Do not make measurement on a circuit above 600VDC This may cause shock hazard or damage to the instrument or equipment under test Do not make measurement with the test leads plugged into the instrument Correct Wrong 1 Set the Function Selector switch to the A position DC should be shown on the upper left corner of the display 2 With the transformer jaws closed and without clamping them o...

Page 4: ... LCD display 3 With the tip of the test leads shorted together press the button to offset the resistance of the test leads 4 Press the button once to enter from the normal mode to the conti nuity check mode should be indicated on the display 5 Connect the tip of the test leads to the circuit under test If the resistance is 20Ω or less the buzzer beeps 2 8 Frequency Measurement On ACA or ACV range ...

Page 5: ...witched from the normal 3 times sec to 2 times sec to reduce fluctuation of the reading LoHz is also indicated where input is 3 of full scale or less 3 4 Optional Accessories Clamp Adapter Model 99025 For AC current measurement only Clamp Adapter Model 99025 is designed to increase the measuring capability of a clamp meter With the use of the Clamp Adapter you can not only extend current range ove...

Page 6: ... The scales are however actually calibrated in terms of the effective value of a sine wave though the clamp meter is responding to the average value The calibration is done with a conversion factor of 1 111 for sine wave which is found by dividing the effective value by the average value These instruments are therefore in error if the input voltage or current has some other shape than sine wave Wa...

Page 7: ...r still fails to operate correctly and needs repair con tact the vendor from whom you purchased the instrument or the nearest Yokogawa M C sales office Turn off the POWER switch once then turn it back on again If the tester does not turn on replace the battery with a new one Calibration It is recommended that the instrument be calibrated once every year ...

Page 8: ...IA PTY LTD AUSTRALIA Phone 61 2 9805 0699 Facsimile 61 2 9888 1844 YOKOGAWA BLUE STAR LTD INDIA Phone 91 80 227 1513 Facsimile 91 80 227 4270 LTD YOKOGAWA ELECTRIC RUSSIAN FEDERATION Phone 7 095 737 7868 Facsimile 7 095 737 7869 計 測 機 器 営 業 部 180 8750 東京都武蔵野市中町 2 9 32 電話 0422 52 5984 ファクシミリ 0422 55 8953 関 西 支 店 564 0063 大阪府吹田市江坂町 1 23 101 大同生命江坂ビル 10 階 電話 06 6368 7041 ファクシミリ 06 6368 7045 中 部 支 店 4...

Reviews: