IM CA71-E
4-15
Source
1
2
3
4
5
6
7
8
9
10
11
12
13
App
4.5.2 Sourcing the Preset Number of Pulses
(Pulse Cycle)
Step 1:
Using the function selector switch, select
.
The LCD shows the default frequency
.
Step 2:
Using the
key, set the frequency range. Each press of
the
key cycles through the 500.0 Hz, 1000 Hz and
10 kHz options.
Step 3:
Set the output value digit by digit using each pair of
and
output setting keys. Each pair of
and
keys
corresponds to each digit of the LCD reading. Each press of
the
or
key increases or decreases the digit.
Increasing the digit from 9 or decreasing it from 0 causes the
digit to overflow or underflow, allowing you to set the output
value without interruption. Holding down the
or
key
continuously changes the digit in question.
Pressing the
key initializes the output setpoint to the
default value (differs depending on the frequency range).
Step 4:
Pressing the
key once switches to amplitude setting
mode. The LCD provides a reading of
.
Step 5:
Set the output value digit by digit using each pair of
and
output setting keys.
Each pair of
and
keys corresponds to each digit of the
LCD reading. Each press of the
or
key increases or
decreases the digit. Increasing the digit from 9 or decreasing
it from 0 causes the digit to overflow or underflow, allowing
you to set the output value without interruption. Holding down
the
or
key continuously changes the digit in question.
Pressing the
key initializes the output setpoint to the de-
fault value (0.1 V).
Step 6:
Press the
key once again to show
on the LCD.
Then, press the
key. The source setpoint reading of the
LCD changes to a numeric value, which represents the
number of pulses.
4.5 Sourcing Pulse Signals
Summary of Contents for CA71 HANDY CAL
Page 110: ......