Calibrating the Pressure T
ransmitter
3-3
IM CA700-02EN
1
2
3
4
5
6
App
Index
3.2 Setting the Device Information and the
Calibration Conditions of the Pressure
Transmitter
Procedure
1.
From the top menu, use the ▲ or ▼ key to select
Calibrate
, and press
ENTER
.
The calibration display appears.
2.
Press
Select Proc. (F2)
.
A list of registered calibration procedures is displayed.
The table lists tag numbers (first 20 characters of each one) and model numbers (first 15 characters of
each one).
Up to 20 sets of procedures can be registered. They are displayed in four windows.
3.
Press or
to change the page and ▲ or ▼ to select the calibration procedure you want to
edit. Press ENTER.
Information of the selected calibration procedure is displayed.
4.
Press
Edit (F1)
.
The calibration procedure setup screen appears. The setup screen is divided into three pages: Transmitter
Information, Measure, and Source.
Hold down ▲ or ▼ to change the screen.