4-22
IM AQ7277-17EN
4.4 Condition Register
The condition register indicates the internal condition of the instrument.
AR
6
0
7
5
4
3
2
1
0
PC
0
AS
0
AVE REAL
0
0
8
0
9
0
10
11
0
12
0
13
WRM
14
0
15
Condition Register
:STATus:CONDition?
The meaning of each bit of the condition register is as follows:
Bit 0
REAL
Set to 1 while the realtime measurement is in progress.
Bit 1 AVE
Set to 1 while the averaging measurement is in progress.
Bit 3 AS
Set to 1 while auto searching.
Bit 5 PC
Set to 1 while the plug is being checked.
Bit 6 AR
Set to 1 while the auto range measurement is in progress.
Bit 14 WRM
Set to 1 during warm-up.