1-11
IM AQ6376-01EN
Functions
1
2
3
4
5
6
7
8
9
10
11
App
Index
1.3 Waveform Display
Split Display
<<See section 5.10 for the operating procedure>>
You can split the screen into an upper and lower display (SPLIT mode).
You can assign trace waveforms to either the upper or lower split.
TRACE A UP/LOW
This function is used to set whether to put trace A on top or on bottom during upper/lower
split display. If you select UP, it is assigned to the top. (Default) If you select LOW, it is
assigned to the bottom.
Noise Mask <<See section 5.11 for the operating procedure>>
This key is used to display a waveform so that parts of the waveform at or below the
set value are masked. In addition, when a noise mask value is set, the waveform is
overwritten in real time.
When you press the NOISE MASK soft key, the current noise mask value is displayed in
the noise mask value setting screen. The allowed settings for the noise mask setting are
OFF (-210 dBm), and the range from -100 to 0 dBm (fine: in steps of 1; coarse: in steps
of 10).
Peak/Bottom Search <<See section 5.13 and 5.14 for the operating procedure>>
The peak/bottom search function has two search modes: single, in which a single peak
or a single bottom of the measured waveform's level are detected, and multi, in which
multiple peaks and bottoms are detected in a single search. Markers are displayed at the
peak and bottom points. You can also detect the next level peak or bottom.
There is also an auto search function that automatically performs peak/bottom searches
during each sweep. This is useful when you want to observe the changes to the peak