23
IM AQ6370D-02EN
Cleaning the Optical Fiber End Face
1.
Firmly press the connector end face of the optical fiber against the cleaning
surface of the cleaner.
2.
While pressing the end face against the cleaner, turn it once.
3.
While pressing the end face against the cleaner, move it.
4.
Repeat steps 1 to 3.
Note
• If you do not firmly press the connector end face of the optical fiber against the cleaner, the
end face may not be cleaned completely.
• You can purchase an optical fiber connector cleaner from NTT-AT Corporation.
Connecting Optical Fibers
5.
Open the instrument’s optical input connector cover.
6.
Connect the optical fiber’s optical connector to the optical input connector on the
instrument.
Setting the optical fiber connector
7.
Press
SETUP
. The soft key menu is displayed.
8.
Press the
MORE 1/2
soft key.
9.
Press the
FIBER CONNECTOR
soft key. Pressing the key repeatedly toggles
between NORM and ANGLED. Set ANGLED if the optical fiber under test is APC
(angle lap PC). Otherwise, set NORMAL.
Note
• If you set FIBER CONNECTOR to ANGLED, is displayed in the
measurement conditions area.
• The instrument's measurement accuracy specification is for when FIBER CONNECTOR is
set to NORM
9 Connecting the DUT