Clock Signal Output Terminals (CLOCK OUT and CLOCK OUT)
These terminals transmit 10 GHz non-inverted and inverted clock signals respectively.
Item
Specifications
Connector type
SMA, female
Output level
0.6 Vp-p (typical) AC-coupling
Offset voltage
–2 to +3 V (in 10 mV steps)
Output terminator conditions 50
Ω
AC or DC terminator
Tr/Tf (20 to 80%)
25 ps or less
Duty
50% ± 10%
Data Input Terminal (DATA IN 1 (CDR))
This receives ED data. It has a built-in CDR.
Item
Specifications
Connector type
3.5 mm, female
Input level
0.1 to 0.7 Vp-p
Minimum input sensitivity
100 mVp-p or less
Threshold voltage
±0.35 V (in 1 mV steps)
Input terminator conditions
50
Ω
AC-coupling
Bitrate
9.95 to 11.32 Gbit/s
Allowable bitrate
±100 ppm of the PPG’s operation bitrate
Data format
NRZ
Data Input Terminal (DATA IN 2)
This terminal receives 10 Gbit/s data.
Item
Specifications
Connector type
3.5 mm, female
Input level
0.1 to 0.6 Vp-p
Minimum input sensitivity
100 mVp-p or less
Threshold voltage
±0.3 V (in 1 mV steps)
Input terminator conditions
50
Ω
AC-coupling
Bitrate
9.95 to 10.71 Gbit/s
Clock Signal Input Terminal (CLOCK IN)
This terminal receives 10 Gbit/s clock signals.
Item
Specifications
Connector type
SMA, female
Input level
0.2 to 0.6 Vp-p
Input terminator conditions
50
Ω
AC-coupling
Frequency
The same frequency as DATA IN 2 (synchronized to the data input)
3.3 BERT Module
3-6
IM 735101-01EN